September 2005 NSF 05/13b/EPADWCTR EPA/600/R-05/108 Environmental Technology Verification Report Removal of Chemical Contaminants in Drinking Water Kinetico Incorporated Pall/Kinetico Purefecta™ Drinking Water Treatment System Prepared by NSF International Under a Cooperative Agreement with U.S. Environmental Protection Agency ------- ------- THE ENVIRONMENTAL TECHNOLOGY VERIFICATION PROGRAM ET VAIVI ^^ V U.S. Environmental Protection Agency NSF International ETV Joint Verification Statement TECHNOLOGY TYPE: POINT-OF-USE DRINKING WATER TREATMENT SYSTEM APPLICATION: REMOVAL OF CHEMICAL CONTAMINANTS IN DRINKING WATER PRODUCT NAME: PALL/KINETICO PUREFECTA™ COMPANY: KINETICO INCORPORATED ADDRESS: 10845 KINSMAN ROAD PHONE: 800-944-9283 NEWBURY, OH 44065 FAX: 440-564-9541 EMAIL: CUSTSERV@KINETICO.COM NSF International (NSF) manages the Drinking Water Systems (DWS) Center under the U.S. Environmental Protection Agency's (EPA) Environmental Technology Verification (ETV) Program. The DWS Center recently evaluated the performance of the Pall/Kinetico Purefecta™ point-of-use (POU) drinking water treatment system. NSF performed all of the testing activities, and also authored the verification report and this verification statement. The verification report contains a comprehensive description of the test. EPA created the ETV Program to facilitate the deployment of innovative or improved environmental technologies through performance verification and dissemination of information. The goal of the ETV Program is to further environmental protection by accelerating the acceptance and use of improved and more cost-effective technologies. ETV seeks to achieve this goal by providing high-quality, peer- reviewed data on technology performance to those involved in the design, distribution, permitting, purchase, and use of environmental technologies. ETV works in partnership with recognized standards and testing organizations, stakeholder groups (consisting of buyers, vendor organizations, and permitters), and with the full participation of individual technology developers. The program evaluates the performance of innovative technologies by developing test plans that are responsive to the needs of stakeholders, conducting field or laboratory tests (as appropriate), collecting and analyzing data, and preparing peer reviewed reports. All evaluations are conducted in accordance with rigorous quality assurance protocols to ensure that data of known and adequate quality are generated and that the results are defensible. NSF 05/13b/EPADWCTR The accompanying notice is an integral part of this verification statement. September 2005 VS-i ------- ABSTRACT The Pall/Kinetico Purefecta™ POU drinking water treatment system was tested for removal of aldicarb, benzene, cadmium, carbofuran, cesium, chloroform, dichlorvos, dicrotophos, fenamiphos, mercury, mevinphos, oxamyl, strontium, and strychnine. The Purefecta™ employs several components: a reverse osmosis (RO) membrane, carbon filters, and a bacteria/virus removal filter to treat drinking water. Treated water is stored in a three-gallon storage tank. The system was first tested with only the RO membrane component in place. The target challenge concentration of each chemical for each RO membrane test was 1 mg/L. Following the RO membrane challenges, the post-membrane carbon filter component was challenged alone with each chemical that the RO membrane did not remove to below 10 Hg/L, except for cesium, which is not well removed by carbon. The target feed concentration of each chemical to a carbon filter component was the maximum effluent level measured during the RO membrane tests. A total of 20 RO membrane components were tested, divided into ten pairs. Each pair of membranes was tested with only one of the ten organic chemicals because of concern that a chemical could compromise the integrity of the membrane material or membrane seals. One pair of RO membrane components was also challenged with the inorganic chemicals. Each RO membrane chemical challenge was conducted over a one-day period. Influent and effluent samples were collected during the operation period, and also the next morning. Post-membrane carbon filter challenges were conducted over a 15-hour duration. Two filters were tested for each chemical challenge, and each pair was only used for one challenge. Influent and effluent samples were collected at the beginning, middle, and end of the challenge period. The Purefecta™ system as a whole, considering both the RO membrane challenge and post-membrane carbon filter challenge results combined, reduced all of the challenge chemicals by 99% or more, except for cesium. TECHNOLOGY DESCRIPTION The following technology description was provided by the manufacturer, and has not been verified. The Purefecta™ is a five-stage POU drinking water treatment system. It employs activated carbon filters and an RO membrane to remove chemical contaminants from drinking water, and a mechanical filtration "biofilter" to remove microorganisms. The system includes a three-gallon maximum capacity pressurized bladder tank for storing the treated water, and a faucet to mount on the kitchen sink. The biofilter is manufactured by the Pall Corporation and supplied to Kinetico, who manufactures the system. The influent water first passes through a pre-membrane sediment or carbon filter, and then through the RO membrane. The permeate water travels through the first stage of the Pall biofilter for virus removal, and then into the storage tank. When the flow of water into the system is started, treated water will be continually produced until the storage tank is nearly full. At that time, the water pressure in the tank activates an automatic shut-off device, stopping the flow of water through the system. After a portion of the water is dispensed from the storage tank, the shut-off device deactivates, allowing water to again flow into the system until the storage tank is nearly full. When the user opens the faucet, the partially treated water exits the storage tank, passes through the post-membrane carbon filter, and finally through the bacteria removal portion of the Pall biofilter. The Purefecta™ is designed to produce approximately four gallons of reject water for every gallon of treated water. NSF 05/13b/EPADWCTR The accompanying notice is an integral part of this verification statement. September 2005 VS-ii ------- VERIFICATION TESTING DESCRIPTION Test Site The testing site was the Drinking Water Treatment Systems Laboratory at NSF in Ann Arbor, Michigan. A description of the test apparatus can be found in the test/QA plan and verification report. The testing was conducted August through November of 2004. Methods and Procedures Verification testing followed the procedures and methods detailed in the Test/QA Plan for Verification Testing of the Pall/Kinetico Purefecta™ Point-of-Use Drinking Water Treatment System for Removal of Chemical Contamination Agents. Because any contamination event would likely be short-lived, the challenge period for each chemical lasted only one day. Long-term performance over the life of the membrane was not investigated. The system was first tested with only the RO membrane component in place. The complete Purefecta™ system, including the storage tank, was used for these tests, but empty cartridges were used in place of the carbon and bacteria/virus filters. A total of 20 RO membranes were challenged with the chemicals in Table 1. The target challenge concentration for each chemical was 1 mg/L. The 20 RO membrane components were divided into ten pairs. Each pair was tested with only one of the ten organic chemicals because of concern that a chemical, especially benzene or chloroform, could compromise the integrity of the membrane material or membrane seals. One pair of the RO membrane components was also challenged with the inorganic chemicals. The inorganic chemical challenges were conducted prior to the organic challenges to eliminate the possibility of damage to the membranes that could bias the inorganic chemical challenge results. Reduction of total dissolved solids (TDS) was also measured to evaluate whether any organic chemicals damaged the membrane material or membrane seals during the challenges. Table 1. Challenge Chemicals Organic Chemicals Inorganic Chemicals Aldicarb Cadmium Chloride Benzene Cesium Chloride (nonradioactive isotope) Carbofuran Mercuric Chloride Chloroform Strontium Chloride (nonradioactive isotope) Dicrotophos Dichlorvos Fenamiphos Mevinphos Oxamyl Strychnine Prior to chemical challenge testing, the RO membrane components were service-conditioned for seven days by feeding the systems test water without any chemical spikes. After completion of the conditioning period, the membranes were subjected to a TDS reduction test using sodium chloride to verify that they were operating properly. Each RO membrane chemical challenge was conducted over a one-day period. The systems were operated for six tank-fill periods, and then were allowed to rest overnight. Influent and effluent samples were collected at start-up, after the 3rd tank fill, after the 5th tank fill, and the next morning after the membranes rested under pressure overnight. NSF 05/13b/EPADWCTR The accompanying notice is an integral part of this verification statement. September 2005 VS-iii ------- Following the RO membrane challenges, post-membrane carbon filters were challenged with the chemicals that the RO membrane did not remove to below 10 ng/L, except for cesium, which is not well removed by carbon. The carbon filters were attached to a separate manifold that was of the same design as the manifold in the full system. The pre-membrane carbon filter was not tested because it is only designed to remove chlorine to protect the RO membrane. Two carbon filter components were tested for each chemical challenge, and each filter was only used for one challenge. The target challenge concentrations were the maximum effluent levels measured during the RO membrane tests. Prior to testing, each carbon filter was service-conditioned by feeding water containing chloroform to simulate the potential chemical loading on the carbon halfway through the filter's effective lifespan. The target chloroform concentration was 300 ± 90 |~ig/L, which is the influent challenge concentration for the VOC reduction test in NSF/ANSI Standard 53 (chloroform is the surrogate challenge chemical for the test). The filters were operated at a flow rate of 0.5 gallons per minute (gpm) for 250 gallons (Kinetico's design capacity for the filter is 500 gallons). The post-membrane carbon filter challenges were 15 hours in duration. Influent and effluent samples were collected at the beginning, middle, and end of the challenge period. The carbon filters were operated on an "on/off operation cycle where the "on" portion was the time required to empty the system storage tank when full, and the "off portion was the time required to fill the storage tank. VERIFICATION OF PERFORMANCE The results of the RO membrane challenges are presented in Table 2. The RO membrane treatment process removed 96% or more of all challenge chemicals except mercury and chloroform. The TDS reduction by each membrane component for all challenge tests was 94% or higher, and the TDS levels in the treated water samples did not increase through any of the challenge periods. This indicates that none of the chemicals compromised the performance of the membrane components to a degree that could be detected. Table 2. RO Membrane Challenge Data Mean Influent Mean Effluent Percent Chemical (|ig/L) (|qg/L) Reduction (%) Cadmium Cesium Mercury Strontium Aldicarb Benzene Carbofuran Chloroform Dichlorvos Dicrotophos Fenamiphos Mevinphos Oxamyl Strychnine 1000 1000 1100 850 950 1100 950 1100 1100 790 740 1400 980 1100 1.9 40 680 2 7 48 6 170 23 ND(10) 2 19 5 18 >99 96 38 99 >99 96 >99 85 98 99 >99 99 >99 98 NSF 05/13b/EPADWCTR The accompanying notice is an integral part of this verification statement. September 2005 VS-iv ------- The post-membrane carbon filter components were challenged with mercury, benzene, chloroform, dichlorvos, mevinphos, and strychnine, based on the criteria that the RO membrane challenge effluents were above 10 |~ig/L. The target challenge levels were the maximum effluent levels measured during the RO membrane challenges. The carbon filters were operated at 0.8 gpm on an operation cycle where the "on" portion was four minutes and thirty seconds, and the "off portion was one hour and ten minutes. The carbon challenge results are shown below in Table 3. The carbon filters reduced all substances to non-detectible levels, except for mercury. However, the mean effluent value for mercury was only 2.7 Hg/L, which still gives a percent reduction greater than 99%. Note that the percent reduction of strychnine was limited by the detection limit for the chemical. The RO membrane and carbon challenge data combined shows that the two treatment technologies working in concert within the Purefecta™ system removed 99% or more of all of the challenge chemicals, except for cesium. Complete descriptions of the verification testing results are included in the verification report. Table 3. Post-Membrane Carbon Filter Challenge Data Mean Influent Mean Effluent Percent Chemical (|qg/L) (ng/L) Reduction (%) Mercury Benzene Chloroform Dichlorvos Mevinphos Strychnine 960 83 320 29 20 31 2.7 ND (0.5) ND (0.5) ND (0.2) ND (0.2) ND(5) >99 >99 >99 >99 99 84 QUALITY ASSURANCE/QUALITY CONTROL (QA/QC) NSF ETV and QA staff monitored the testing activities to ensure that the testing was in compliance with the test plan. NSF also conducted a data quality audit of 100% of the data. Please see the verification report referenced below for more QA/QC information. NSF 05/13b/EPADWCTR The accompanying notice is an integral part of this verification statement. September 2005 VS-v ------- Original signed by Andrew Avel, 9/26/05 Original signed by Robert Ferguson, 10/5/05 Andrew P. Avel Date Robert Ferguson Date Acting Director Vice President National Homeland Security Research Center Water Systems United States Environmental Protection NSF International Agency NOTICE: Verifications are based on an evaluation of technology performance under specific, predetermined criteria and the appropriate quality assurance procedures. EPA and NSF make no expressed or implied warranties as to the performance of the technology and do not certify that a technology will always operate as verified. The end-user is solely responsible for complying with any and all applicable federal, state, and local requirements. Mention of corporate names, trade names, or commercial products does not constitute endorsement or recommendation for use of specific products. This report is not an NSF Certification of the specific product mentioned herein. Availability of Supporting Documents Copies of the test protocol, the verification statement, and the verification report (NSF report # NSF 04/13b/EPADWCTR) are available from the following sources: (NOTE: Not all of the appendices are included in the verification report. The appendices are available from NSF upon request.) 1. ETV Drinking Water Systems Center Manager (order hard copy) NSF International P.O. Box 130140 Ann Arbor, Michigan 48113-0140 2. NSF web site: http://www.nsf.org/etv/dws/dws_reports.html, and from http://www.nsf.org/etv/dws/dws_project_documents.html (electronic copy) EPA web site: https://www.epa.gov/etv (electronic copy) NSF 05/13b/EPADWCTR The accompanying notice is an integral part of this verification statement. September 2005 VS-vi ------- September 2005 Environmental Technology Verification Report Removal of Chemical Contaminants in Drinking Water Kinetico Incorporated TM Pall/Kinetico Purefecta Drinking Water Treatment System Prepared by: NSF International Ann Arbor, Michigan 48105 Under a cooperative agreement with the U.S. Environmental Protection Agency Jeffrey Q. Adams, Project Officer National Risk Management Research Laboratory U.S. Environmental Protection Agency Cincinnati, Ohio 45268 ------- Notice The U.S. Environmental Protection Agency (USEPA) through its Office of Research and Development has financially supported and collaborated with NSF International (NSF) under Cooperative Assistance Agreement No. R-82833301. This verification effort was supported by the Drinking Water Systems (DWS) Center, operating under the Environmental Technology Verification (ETV) Program. This document has been peer-reviewed, reviewed by NSF and USEPA, and recommended for public release. ------- Foreword The U.S. Environmental Protection Agency (USEPA) is charged by Congress with protecting the Nation's land, air, and water resources. Under a mandate of national environmental laws, the Agency strives to formulate and implement actions leading to a compatible balance between human activities and the ability of natural systems to support and nurture life. To meet this mandate, USEPA's research program is providing data and technical support for solving environmental problems today and building a science knowledge base necessary to manage our ecological resources wisely, understand how pollutants affect our health, and prevent or reduce environmental risks in the future. The National Risk Management Research Laboratory (NRMRL) is the Agency's center for investigation of technological and management approaches for preventing and reducing risks from pollution that threaten human health and the environment. The focus of the Laboratory's research program is on methods and their cost-effectiveness for prevention and control of pollution to air, land, water, and subsurface resources; protection of water quality in public water systems; remediation of contaminated sites, sediments and ground water; prevention and control of indoor air pollution; and restoration of ecosystems. NRMRL collaborates with both public and private sector partners to foster technologies that reduce the cost of compliance and to anticipate emerging problems. NRMRL's research provides solutions to environmental problems by: developing and promoting technologies that protect and improve the environment; advancing scientific and engineering information to support regulatory and policy decisions; and providing the technical support and information transfer to ensure implementation of environmental regulations and strategies at the national, state, and community levels. This publication has been produced as part of the Laboratory's strategic long-term research plan. It is published and made available by USEPA's Office of Research and Development to assist the user community and to link researchers with their clients. Sally Gutierrez, Director National Risk Management Research Laboratory in ------- Table of Contents Verification Statement VS-i Title Page i Notice ii Foreword iii Table of Contents iv List of Tables vi List of Figures vi Abbreviations and Acronyms vii Acknowlegements viii Chapter 1 Introduction 1 1.1 Environmental Technology Verification (ETV) Program Purpose and Operation 1 1.2 Purpose of Verification 1 1.3 Development of Test/Quality Assurance (QA) Plan 1 1.4 Challenge Chemicals 2 1.5 Testing Participants and Responsibilities 2 1.5.1 NSF International 2 1.5.2 Kinetico Inc 3 1.5.3 Pall Corporation 3 1.5.4 U.S. Environmental Protection Agency 3 Chapter 2 Equipment Description 4 2.1 Principal of Operation 4 2.1.1 RO Membrane 4 2.1.2 Activated Carbon Filtration 4 2.2 Equipment Capabilities 4 2.3 System Components 4 2.4 System Operation 5 2.5 Rate of Waste Production 6 2.6 Equipment Operation Limitations 7 2.7 Operation and Maintenance Requirements 7 Chapter 3 Methods and Procedures 8 3.1 Introduction 8 3.1.1 RO Membrane Challenges 8 3.1.2 Post-Membrane Carbon Filter Challenges 9 3.1.3 System Operation Scenarios 9 3.2 Verification Test Procedure 10 3.2.1 Challenge Protocol Tasks 10 3.2.2 Test Rig 10 3.2.3 Test Water 10 3.2.3.1 RO Membrane Conditioning and Challenge Test Water 10 3.2.3.2 Post-Membrane Carbon Filter Conditioning and Challenge Test Water 11 iv ------- 3.2.3.3 Chemical Challenges 12 3.2.4 Test Unit Installation and Conditioning 12 3.2.4.1 RO Membrane Test Units 12 3.2.4.2 Post-Membrane Carbon Filter Test Units 13 3.2.5 Challenge Protocols and Sampling Plans 13 3.2.5.1 TDS Reduction System Performance Check 13 3.2.5.2 RO Membrane Challenge Testing 13 3.2.5.3 Post-Membrane Carbon Filter Challenge Testing 16 3.3 Analytical Methods 17 3.3.1 Water Quality Analytical Methods 17 3.3.2 Challenge Chemical Analytical Methods 18 Chapter 4 Results and Discussion 19 4.1 RO Membrane Conditioning 19 4.1.1 RO Membrane Test System Operation Data 19 4.2 Post-Membrane Carbon Filter Conditioning 19 4.3 TDS Reduction System Performance Check 19 4.4 RO Membrane Chemical Challenges 20 4.4.1 Inorganic Chemicals Challenge 20 4.4.2 Organic Chemical Challenges 21 4.5 Post-Membrane Carbon Filter Challenges 23 4.6 Conclusions 24 Chapters QA/QC 25 5.1 Introduction 25 5.2 Test Procedure QA/QC 25 5.3 Sample Handling 25 5.4 Analytical Methods QA/QC 25 5.5 Documentation 26 5.6 Data Review 26 5.7 Data Quality Indicators 26 5.7.1 Representativeness 26 5.7.2 Accuracy 26 5.7.3 Precision 27 5.7.4 Completeness 27 5.7.4.1 Number of Systems Tested 28 5.7.4.2 Water Chemistry Measurements 28 5.7.4.3 Challenge Chemicals 28 Chapter 6 References 30 ------- Appendix Appendix A. Conditioning and Chemical Challenges Data Tables 31 List of Tables Table 1-1. Challenge Chemicals 2 Table 3-1. Challenge Chemicals 9 Table 3-2. Summary of Sampling Plan for RO Membrane Challenges 15 Table 3-3. Summary of Sampling Plan for Post-Membrane Carbon Filter Challenges 17 Table 3-4. QC Limits and Method Reporting Limits for Analyses 17 Table 4-1. RO Membrane Test Unit Operation Data 19 Table 4-2. RO Membrane Inorganic Chemicals Reduction Data 20 Table 4-3. Inorganic Chemicals Challenge Reject Water Data 21 Table 4-4. RO Membrane Organic Chemical Challenge Data 22 Table 4-5. TDS Reduction Data for Organic Chemical Challenges 23 Table 4-6. Organic Chemical Challenge Reject Water Data 23 Table 4-7. Post-Membrane Carbon Filter Challenge Data 24 Table 5-1. Completeness Requirements 27 List of Figures Figure 2-1. Cutaway Schematic Diagram of Purefecta™ Treatment Elements 5 Figure 2-2. Photograph of the Purefecta™ 6 Figure 3-1. RO Membrane Systems Installed at a Test Station 14 Figure 3-2. Post-Membrane Carbon Filters Installed at a Test Station 16 VI ------- Abbreviations and Acronyms ANSI °C DWS ETV °F GC/MS gpd gpm HC1 HPLC ICP/MS L mg mL NaOH ND NRMRL NSF NTU POE POU psi QA QC QA/QC RO RPD RSD SOP IDS TOC Mg USEPA American National Standards Institute Degrees Celsius Drinking Water Systems Environmental Technology Verification Degrees Fahrenheit Gas Chromatography/Mass Spectrometry Gallons Per Day Gallons Per Minute Hydrochloric Acid High Pressure Liquid Chromatography Inductively Coupled Plasma - Mass Spectrometry Liter Milligram Milliliter Sodium Hydroxide Non-detect National Risk Management Research Laboratory NSF International (formerly known as National Sanitation Foundation) Nephelometric Turbidity Unit Point-of-Entry Point-of-Use Pounds per Square Inch Quality Assurance Quality Control Quality Assurance/Quality Control Reverse Osmosis Relative Percent Difference Relative Standard Deviation Standard Operating Procedure Total Dissolved Solids Total Organic Carbon Microgram U. S. Environmental Protection Agency vn ------- Acknowledgments NSF was responsible for all elements in the testing sequence, including collection of samples, calibration and verification of instruments, data collection and analysis, data management, data interpretation and the preparation of this report. The manufacturer of the equipment was: Kinetico Incorporated 10845 Kinsman Road Newbury, OH 44065 NSF wishes to thank the members of the expert technical panel for their assistance with development of the test plan. Vlll ------- Chapter 1 Introduction 1.1 Environmental Technology Verification (ETV) Program Purpose and Operation The U.S. Environmental Protection Agency (USEPA) has created the ETV Program to facilitate the deployment of innovative or improved environmental technologies through performance verification and dissemination of information. The goal of the ETV Program is to further environmental protection by accelerating the acceptance and use of improved and more cost- effective technologies. ETV seeks to achieve this goal by providing high-quality, peer-reviewed data on technology performance to those involved in the design, distribution, permitting, purchase, and use of environmental technologies. ETV works in partnership with recognized standards and testing organizations; with stakeholder groups consisting of buyers, vendor organizations, and permitters; and with the full participation of individual technology developers. The program evaluates the performance of innovative technologies by developing test plans that are responsive to the needs of stakeholders, by conducting field or laboratory testing, collecting and analyzing data, and by preparing peer- reviewed reports. All evaluations are conducted in accordance with rigorous quality assurance protocols to ensure that data of known and adequate quality are generated and that the results are defensible. The USEPA has partnered with NSF International (NSF) under the ETV Drinking Water Systems (DWS) Center to verify performance of drinking water treatment systems that benefit the public and small communities. It is important to note that verification of the equipment does not mean the equipment is "certified" by NSF or "accepted" by USEPA. Rather, it recognizes that the performance of the equipment has been determined and verified by these organizations under conditions specified in ETV protocols and test plans. 1.2 Purpose of Verification The purpose of this verification was to evaluate treatment system performance under a simulated intentional or non-intentional chemical contamination event. Because any contamination event would likely be short-lived, the challenge period for each chemical lasted only one day. Long- term performance over the life of the membrane was not investigated. 1.3 Development of Test/Quality Assurance (QA) Plan USEPA's "Water Security Research and Technical Support Action Plan" (USEPA, 2004) identifies the need to evaluate point-of-use (POU) and point-of-entry (POE) treatment system capabilities for removing likely contaminants from drinking water. As part of the ETV program, NSF developed a test/QA plan for evaluating POU reverse osmosis (RO) drinking water treatment systems for removal of chemical contaminants. To assist in this endeavor, NSF ------- assembled an expert technical panel, which gave suggestions on a protocol design prior to development of the test/QA plan. The product-specific test/QA plan for evaluating the Pall/Kinetico Purefecta™ was entitled Test/QA Plan for Verification Testing of the Pall/Kinetico Purefecta™ Point-of-Use Drinking Water Treatment System for Removal of Chemical Contamination Agents. By participating in this ETV evaluation, the vendor obtains USEPA and NSF verified independent test data indicating potential user protection against intentional or non-intentional chemical contamination of drinking water. Verifications following an approved test/QA plan serve to notify the public of the possible level of protection against chemical contamination agents afforded to them by the use of a verified system. 1.4 Challenge Chemicals The challenge chemicals for this verification are listed in Table 1-1. Table 1-1. Challenge Chemicals Organic Chemicals Inorganic Chemicals Aldicarb Cadmium Chloride Benzene Cesium Chloride (nonradioactive isotope) Carbofuran Mercuric Chloride Chloroform Strontium Chloride (nonradioactive isotope) Dicrotophos Dichlorvos Fenamiphos Mevinphos Oxamyl Strychnine 1.5 Testing Participants and Responsibilities The ETV testing of the Pall/Kinetico Purefecta™ was a cooperative effort between the following participants: NSF Kinetico, Inc. USEPA The following is a brief description of each of the ETV participants and their roles and responsibilities. 1.5.1 NSF International NSF is a not-for-profit organization dedicated to public health and safety, and to protection of the environment. Founded in 1946 and located in Ann Arbor, Michigan, NSF has been instrumental ------- in the development of consensus standards for the protection of public health and the environment. The USEPA partnered with NSF to verify the performance of drinking water treatment systems through the USEPA's ETV Program. NSF performed all verification testing activities at its Ann Arbor location. NSF prepared the test/QA plan, performed all testing, managed, evaluated, interpreted, and reported on the data generated by the testing, and reported on the performance of the technology. Contact Information: NSF International 789 N. Dixboro Road Ann Arbor, MI 48105 Phone:734-769-8010 Fax: 734-769-0109 Contact: Bruce Bartley, ETV Program Manager Email: bartley@nsf.org 1.5.2 Kinetico Inc. The verified system is manufactured by Kinetico Inc., a manufacturer of water treatment products for home and business. The manufacturer was responsible for supplying the RO systems in accordance with section 3.1.1, and for providing logistical and technical support as needed. Contact Information: Kinetico Inc. 10845 Kinsman Road Newbury, OH 44065 Phone: 1-800-944-9283 Contact Person: Mr. Rod Yoder Email: ryoder@kinetico.com 1.5.3 Pall Corporation The Pall Corporation is a manufacturer of products for fluid filtration, separation, and purification. Pall manufactures and supplies to Kinetico the bacteria and virus removal filter component for the Purefecta™. 1.5.4 U.S. Environmental Protection Agency The USEPA, through its Office of Research and Development, has financially supported and collaborated with NSF under Cooperative Agreement No. R-82833301. This verification effort was supported by the DWS Center operating under the ETV Program. This document has been peer-reviewed, reviewed by the USEPA, and recommended for public release. ------- Chapter 2 Equipment Description 2.1 Principal of Operation 2.1.1 RO Membrane Membrane technologies are among the most versatile water treatment processes because of their ability to effectively remove a wide variety of contaminants. RO membranes operate by the principal of cross-flow filtration. In this process, the influent water flows over and parallel to the filter medium and exits the system as reject water. Under pressure, a portion of the water diffuses through the membrane becoming "permeate". The membrane allows water molecules to pass through its pores, but not most dissolved inorganic chemical molecules and larger molecular weight organic chemical molecules. These molecules are concentrated in and washed away with the reject water stream. 2.1.2 Activated Carbon Filtration Activated carbon removes organic chemicals from water through the process of adsorption. The chemicals are attracted to and attach to the surface of the carbon through electrostatic interactions. The adsorbent properties of activated carbon are a function of the raw material used and the activation process. Once the carbon is saturated with adsorbed molecules, it must be replaced. 2.2 Equipment Capabilities The Purefecta™ is certified by NSF to NSF/ANSI Standard 58 -Reverse Osmosis Drinking Water Treatment Systems. The system has a certified production rate of 19.8 gallons per day (gpd), and an efficiency rating of 25.9%. Efficiency rating as defined by Standard 58 is "a percentage measure of the amount of influent water that is delivered as permeate under a closed permeate discharge set of actual use conditions." These measurements are based on system operation at 50 pounds per square inch (psi) inlet pressure, a water temperature of 77 °F, and a total dissolved solids (TDS) level of 750 ± 40 mg/L. The amount and quality of treated water produced varies depending on the inlet pressure, water temperature, and level of TDS. These measurements were not subject to verification during this study. 2.3 System Components The Purefecta™ is a five-stage POU treatment system. It uses activated carbon, an RO membrane, UF membranes, and optional sediment filtration to treat drinking water. The system also includes a three gallon maximum capacity storage tank, and a faucet to dispense the treated water. A cutaway diagram of the treatment components is shown in Figure 2-1, and a photograph of the full system is shown in Figure 2-2. The inlet water first passes through an activated carbon or sediment filter, and then through the RO membrane. The permeate water travels through the first stage of the Pall biofilter for virus removal. The partially treated water is ------- then sent to the storage tank. When the user opens the faucet, the water exits the storage tank, passes through the post-membrane activated carbon filter, and then lastly through the bacteria removal portion of the Pall biofilter, before exiting the faucet. Please note that this description, and the system operation description in section 2.4 are given for informational purposes only. This information was not subject to verification. Figure 2-1. Cutaway Schematic Diagram of Purefecta™ Treatment Elements . t Five Stages of Filtration 1. Sediment or Carbon Prefilter 2. Reverse Osmosis Membrane * Removal of dissolved solids such as lead and nitrates 3. Pall Virus Filter DV 20, 0.02 urn " Four log reduction of viruses such as Rotavirus and Norwalk virus Storage Tank (not pictured) A. Carbon Block o Reduces volatile organic chemicals and chlorine 5. Pall Ultipor, 0.2 urn Final Filter » Six log reduction of bacteria such as Klebsietts terigena and Legionella pneumophiSa Independent Faucet 2.4 System Operation When the flow of water into the system is started, treated water will be continually produced until the storage tank is nearly full. At that time, the water pressure in the tank causes an automatic shut-off device to activate, stopping the flow of water through the system. After approximately two-thirds of the water is dispensed from the storage tank, the shut-off device deactivates, allowing water to again flow into the system until the storage tank is nearly full. The operational storage tank capacity will vary slightly from unit to unit, and is also affected by the inlet water pressure, but is approximately two gallons under normal use conditions. The Purefecta™ uses Kinetico's "PureMometer™" filter life indicator to tell the user how much capacity remains on the carbon or sediment prefilter, the biofilter, and the post-membrane carbon ------- filter. The PureMometer™ is located on the top of the system manifold, easily visible to the user (see Figure 2.2). The PureMometer™ indicator stick decreases in height as treated water is produced. After approximately 500 gallons are produced, the system shuts off. The meter is reset when the user replaces the post-membrane carbon filter. Figure 2-2. Photograph of the Purefecta TM 2.5 Rate of Waste Production As discussed in section 2.2, the efficiency rating of the Purefecta™ is 25.9%, which means the system produces approximately three gallons of reject water for each gallon of product water produced. The efficiency rating was not verified as part of this evaluation. ------- 2.6 Equipment Operation Limitations Kinetico gives the following limitations for the drinking water to be treated by the system: • temperature of 35 - 100 °F; • inlet pressure of 40 - 100 psi; • pHof3-ll; • maximum TDS level of 3000 mg/L; • hardness less than 10 grains per gallon; and • iron less than 0.1 mg/L. 2.7 Operation and Maintenance Requirements Kinetico recommends that all maintenance be done by qualified Kinetico professionals. The following are the operation and maintenance requirements: • Replacement of the pre-membrane carbon or sediment filter, the biofilter, and the post- membrane carbon filter annually, or when the meter described in section 2.4 stops the production of treated water; • Sanitization of the system when the carbon filters and biofilter are replaced; and • Measurement of the TDS level of the product water when the other filters are changed. RO membrane replacement is recommended when the TDS reduction performance falls below 90%. ------- Chapter 3 Methods and Procedures 3.1 Introduction The challenge tests followed the procedures described in the Test/QA Plan for Verification Testing of the Pall/Kinetico Purefecta™ Point-of-Use Drinking Water Treatment System for Removal of Chemical Contamination Agents. As described in section 2.3, the Purefecta™ employs an RO membrane, activated carbon filters, and a bacteria/virus removal filter to treat drinking water. The system was first tested with only the RO membrane component in place. After the RO membrane challenges were complete, the post-membrane carbon filter was challenged alone. This approach allowed an evaluation of the individual performance of each component, and also served to simulate a worst-case scenario where the carbon filters are at or past the end of their useful life. This approach also allowed each treatment component to be challenged using a test water that presented more of a worse- case scenario for that component. The pre-membrane carbon filter was not tested, because it is not a standard component, and it is only designed to remove chlorine to protect the RO membrane. The pre-membrane sediment filter also was not tested, because it also is not a standard component, and it is only designed to remove suspended sediment particles to protect the RO membrane from clogging. 3.1.1 RO Membrane Challenges The RO membranes were challenged with each chemical in Table 3-1. The target challenge concentration for each chemical was 1 mg/L, which is much higher than most challenge levels in the NSF/ANSI Standards for POU devices. Only two membranes were challenged with each chemical. The organic chemical challenges were conducted one chemical at a time, but the inorganic chemicals were combined into one challenge. Each membrane was only tested with one of the ten organic chemicals, because of concern that some of them, especially benzene and chloroform, could damage the membranes or membrane seals at the high challenge levels. This approach eliminated the possibility that membrane performance against subsequent chemicals was negatively biased. TDS reduction was also measured during the challenges, to serve as a membrane performance benchmark, and also to evaluate whether any organic chemicals damaged the membrane or integrity of the membrane seals. A total of twenty RO membrane units were tested, divided into ten pairs. The inorganic chemicals challenge was conducted first. The two systems used for the inorganic chemicals challenge were used again for an organic chemical challenge. As discussed in section 1.2, each challenge period was only one day. The systems were operated for six tank-fill periods, and then were allowed to rest overnight. Influent and effluent samples were collected during the operation period, and also the next morning after the rest period. In addition to influent and effluent samples, reject water samples were also collected and analyzed ------- in an attempt to determine whether any of the chemicals adsorbed onto or absorbed into the membrane material in significant amounts. See section 3.2.5.2 for RO membrane challenge protocol details. Table 3-1. Challenge Chemicals Organic Chemicals Inorganic Chemicals Aldicarb Cadmium Chloride Benzene Cesium Chloride (nonradioactive isotope) Carbofuran Mercuric Chloride Chloroform Strontium Chloride (nonradioactive isotope) Dicrotophos Dichlorvos Fenamiphos Mevinphos Oxamyl Strychnine 3.1.2 Post-Membrane Carbon Filter Challenges The post-membrane carbon filter was tested alone with the organic chemicals the RO membrane did not remove to below 10 |ag/L. The inorganic chemicals were considered on a case-by-case basis, since USEPA does not consider carbon to be the best available technology for removing cadmium, cesium, or strontium. As with the membranes, the carbon filters were challenged in pairs, and each pair was only tested once. Each challenge was 15 hours. The target challenge concentrations for the carbon filter tests were the maximum effluent levels measured during the RO tests. See section 3.2.5.3 for the post-membrane carbon filter test protocol details. 3.1.3 System Operation Scenarios The challenge protocol was designed to evaluate system performance under two different operation scenarios. The first is operation with the product water storage tank over half full, giving high back-pressure. This is how the system is likely to operate in the home, as the user will usually dispense small volumes of water until the shut-off valve deactivates, allowing the storage tank to fill again. RO membrane performance is affected by the net driving pressure on the membrane. The net driving pressure is the feed water pressure minus the osmotic pressure minus the back-pressure from the storage tank. As the storage tank fills up and the tank bladder expands, the back-pressure increases, reducing the net driving pressure. As the net driving pressure drops, the ion rejection performance of the membrane can also drop (Slovak, 2000). This test protocol was designed so that the test units operate for multiple tank fills under conditions where the net driving pressure was as low as possible. After the first tank fill, the lab technician dispensed the product water to the drain until the shut-off valve deactivated, allowing the unit to again produce treated water. This cycle was repeated for a total of five storage tank fill periods. ------- The second operation scenario is continued contaminant rejection while the system is at rest. The NSF/ANSI Standard 58 testing protocols call for a two-day stagnation period to check whether the membrane can maintain rejection of the contaminants. NSF has observed that RO systems can give higher contaminant concentrations after the rest period than before. This phenomenon is due to the membrane's difficulty maintaining the osmotic differential across the membrane, and perhaps also imperfections in the membrane material. At the end of each challenge, the test units were allowed to rest under pressure overnight, and product water samples were collected for analysis the next morning. 3.2 Verification Test Procedure 3.2.1 Challenge Protocol Tasks The following are the tasks in the challenge protocol, and the order in which they were conducted: 1. Installation of the RO test units on the test rig, and seven days of conditioning (section 3.2.4.1); 2. One-day TDS challenge test to evaluate system integrity (section 3.2.5.1); 3. Conditioning of the post-membrane carbon filters while the RO membrane tests are being conducted (section 3.2.4.2); and 4. Chemical challenge tests a. RO inorganic chemicals challenge (section 3.2.5.2) b. RO organic chemical challenges (section 3.2.5.2) c. Post-membrane carbon filter challenges (section 3.2.5.3). 3.2.2 Test Rig All test units were plumbed to "injection rig" test stations in the NSF Drinking Water Treatment Systems Testing Laboratory. The injection rigs have a common 90-gallon tank to hold the test water without the challenge chemicals. Fresh water is periodically added to the tank as it is being used. Online monitors and a computer system automatically control the water level and water chemistry. Downstream of the feedwater tank a precisely controlled pump is used to inject the challenge chemical(s) at the proper concentrations. Immediately downstream of the pump lies a motionless in-line mixer to assure complete mixing of the challenge water. An influent sample port is downstream of the in-line mixer. No schematic diagram of the injection rig is available, due to the proprietary nature of the design. 3.2.3 Test Water 3.2.3.1 RO Membrane Conditioning and Challenge Test Water The test water for the RO membrane conditioning and challenges was a synthetic water constructed from deionized municipal drinking water. The municipal water was first filtered through activated carbon to remove chlorine, then it was deionized and treated with reverse osmosis. Sodium chloride was added for TDS, and the pH was adjusted with hydrochloric acid 10 ------- (HC1) or sodium hydroxide (NaOH), if necessary, to achieve the following characteristics prior to addition of the challenge chemical(s): • pH - 7.5 ± 0.5 for the TDS reduction test, conditioning, and organic chemical challenges, 6.0-6.5 for the inorganic chemicals challenge; • total chlorine - < 0.05 mg/L; • temperature - 25 ± 1 °C; • TDS - 750 ± 75 mg/L; and • turbidity - < 1 Nephelometric Turbidity Unit (NTU). TDS, pH, temperature, and turbidity were maintained within the appropriate range by a computer system with on-line monitors. In addition, grab samples were collected and analyzed for all parameters according to the sampling plans described in sections 3.2.4.1, 3.2.5.1, and 3.2.5.2. Note that the pH specification for the inorganic chemicals challenges was 6.0 to 6.5, to ensure that the inorganic chemicals were present as dissolved free ions in the challenge water. This ensured that the inorganic chemicals challenges were testing the ability of the RO membrane to reject the ions instead of physically removing suspended particles of the inorganic chemicals. 3.2.3.2 Post-Membrane Carbon Filter Conditioning and Challenge Test Water The test water for post-membrane carbon filter conditioning and testing was the "general test water" specified in NSF/ANSI Standard 53, Drinking water treatment units - health effects (NSF International, 2002). This water is the Ann Arbor municipal drinking water that is adjusted, if necessary, to have the following characteristics prior to addition of the challenge chemical: • pH-7.5 ±0.5; • TDS - 200-500 mg/L • temperature - 20 ± 2.5 °C; • total organic carbon (TOC) - > 1.0 mg/L; and • turbidity - < 1 NTU. Please note that the TOC parameter only has a minimum level specified, since it is the natural TOC in the municipal water supply. During testing, the TOC in the water supply ranged from 1.9 to 2.6 mg/L. However, the TOC levels in the organic chemical challenge waters were much higher due to the methanol used as the carrier solution for the chemicals. TDS, pH, and temperature were maintained within the appropriate range by a computer system with on-line monitors. The pH of the Ann Arbor drinking water was above 7.5 during the test period, so the pH was adjusted with HC1. The TDS level was within the allowable range, so no adjustments were needed. The water was not dechlorinated prior to use. Grab samples were collected and analyzed for all parameters according to the sampling plans described in sections 3.2.4.2 and 3.2.5.3. Total chlorine was also measured, although there is no specification given for it as there is in section 3.2.3.1 for the RO membrane test water. 11 ------- 3.2.3.3 Chemical Challenges The appropriate chemical(s) were added to the base test waters given in sections 3.2.3.1 and 3.2.3.2 to make the challenge waters. The RO membrane challenge target concentration for each chemical was 1 ± 0.5 mg/L. The target challenge concentrations for the carbon filter tests were the maximum effluent levels measured during the RO tests. For each challenge, concentrated solution of the chemical(s) was made, and this mixture injected into the influent water stream at an appropriate rate. Due to analytical procedure lengths, the amount of chemical to add to the test water to achieve the proper challenge concentration was calculated based on the known concentration in the feed solution. The tests were conducted without waiting for confirmation of the influent level from the chemistry laboratory. All challenge chemical influent samples were within the allowable limits. 3.2.4 Test Unit Installation and Conditioning 3.2.4.1 RO Membrane Test Units The RO membrane test units were delivered and installed on the test rigs by Kinetico representatives. Kinetico's recommended conditioning procedure of operation for two tank-fill periods was not conducted, instead the units underwent a seven day, seven tank-fills conditioning period. Previous POU RO system ETV tests for microbial agents indicated that perhaps membrane performance does not stabilize until after four or five days (four or five tank fills) of conditioning. A seven-day conditioning period ensured that the membranes were performing optimally prior to the chemical challenges. For the first six days, the units were operated at 60 ± 3 psi inlet pressure for one storage tank fill period per day using the water described in section 3.2.3.1. Influent water samples were collected each day at the beginning of the operation period for analysis of pH, TDS, temperature, total chlorine, and turbidity. The units rested under pressure overnight, and the storage tanks were emptied the next morning prior to beginning that day's operation period. On the seventh day, the units were instead operated at 80 ± 3 psi inlet pressure. Influent water samples were collected at the beginning of the operation period for analysis of pH, TDS, temperature, total chlorine, and turbidity. The times required to fill the storage tanks were measured and recorded for the three units whose tanks filled the fastest. On the morning of the eighth day, the times to dispense the first liter of water and to empty the storage tanks with the faucet fully open were measured and recorded for the three units whose operating times were recorded the previous day. The tank fill times, times to empty the storage tank, and first liter flow rate data were used to determine the operating parameters for the post-membrane carbon filters during the carbon filter challenge tests. The longest time to empty the storage tank was used for the "on" time portion of the operation cycle. The shortest tank fill time was used for the "off portion of the cycle. The flow rates for the carbon filter challenges were set at the fastest first liter flow rate. Operation at 80 psi instead of 60 psi caused the tank fill time to be shorter, which gave a worse case testing scenario for the carbon filters. See section 3.2.5.3 for further discussion about the post-membrane carbon filter challenge tests. 12 ------- 3.2.4.2 Post-Membrane Carbon Filter Test Units The carbon filters were plumbed to a test station and operated using the water described in section 3.2.3.2 amended with 300 ± 90 |~ig/L of chloroform until 250 gallons passed through each filter. This is the volume equal to one-half of Kinetico's stated capacity of 500 gallons for the filter. The filters were operated at an inlet water pressure of 60 ± 3 psi and a maximum flow rate of 0.5 gallons per minute (gpm), on a ten minutes on, ten minutes off cycle. Chloroform at 300 |ag/L is the influent challenge concentration for the organic chemical reduction test in NSF/ANSI Standard 53 (chloroform is the surrogate chemical for the Standard 53 organic chemical reduction claim). The chloroform served to load the carbon filters to a degree that simulated contaminant loading in the middle of their effective lifespan. Influent samples were collected for analysis of chloroform, pH, temperature, TOC, and turbidity at start-up, approximately 25% of capacity, and approximately 50% of capacity. Effluent samples were collected at the same three points for chloroform analysis. If the filters were not immediately used for a challenge test, they were stored with the conditioning water still in them. The manifold inlets and outlets were closed off by valves to ensure that all of the chloroform remained on the carbon. 3.2.5 Challenge Protocols and Sampling Plans 3.2.5.1 TDS Reduction System Performance Check After the RO membrane conditioning period was complete, they underwent a short-term TDS reduction test to verify that they were operating properly. The challenge was conducted as follows: 1. The product water storage tanks were drained, and test unit operation was started at 50 ± 3 psi inlet pressure using the water described in section 3.2.3.1 without any challenge chemicals added. 2. Immediately after the units began operation, influent samples were collected for analysis of pH, temperature, total chlorine, turbidity, and TDS. 3. The systems were allowed to operate until the automatic shut-off mechanisms activated. 4. The entire contents of the storage tanks were emptied into separate containers, and three 250 mL samples were collected from each container for TDS analysis. Removal of 75% or more of the TDS was required for the use of each membrane for the chemical challenges. 3.2.5.2 RO Membrane Challenge Testing As discussed in section 3.1.1, the RO membrane systems were divided into ten pairs. The inorganic chemical challenges were conducted first, followed by the organic chemicals. Figure 3-1 shows a pair of test units plumbed to the test rig. The challenge tests were conducted as follows: 1. At the start of each challenge period, the system storage tanks were emptied. 13 ------- 2. The initial dynamic inlet water pressure was set at 50 ± 3 psi, and system operation was started using the test water described in section 3.2.3.1 with the proper challenge chemical(s) added. 3. Influent and effluent water samples were collected for analysis of the challenge chemical(s) and TDS immediately after the units began operation. Influent samples were also collected for analysis of pH, temperature, total chlorine, and turbidity. All influent and effluent samples for challenge chemical analysis were collected and analyzed in triplicate, except where indicated below. To collect the triplicate samples, the volumes necessary to obtain the triplicate samples were first collected into a polyethylene container, and then the triplicate samples were collected from that volume. However, due to the volatility of benzene and chloroform, true triplicate samples were not collected for these chemicals. Instead, three consecutive replicate samples were collected directly into the sample bottles that were delivered to the NSF Chemistry Laboratory. TDS samples were collected as single samples. Figure 3-1. RO Membrane Systems Installed at a Test Station 14 ------- 4. While under operation for the first storage tank fill period, duplicate samples were collected from the reject water line of one of the test units for challenge chemical(s) analysis at start- up, approximately halfway through, and approximately three-fourths of the way through the period. 5. The units were operated continuously until the shut-off valves activated. The faucets were then fully opened, and a minimum of one liter, the volume required for sample analysis, or the amount needed to fully deactivate the shut-off valve, was dispensed to drain from each system. Full deactivation was estimated by monitoring resumption of the flow of reject water as the product water was dispensed. The shut-off valve was considered fully deactivated when the flow of reject water appeared to have fully resumed. 6. Step 5 was repeated until five storage tank fill periods were complete. After the third storage tank fill period ended, influent and effluent samples were collected for analysis of the challenge chemical(s) and TDS. 7. Approximately halfway through the last tank fill period, duplicate reject water samples were collected for challenge chemical(s) analysis. The samples were collected from the same system from which the reject water samples were collected in step 4. This sample served to check whether any chemical adsorption/absorption observed during the first storage tank fill period was still occurring, or the membrane became saturated with the chemical. 8. After the fifth storage tank fill, effluent samples were collected from each system for challenge chemical(s) and TDS analysis. Influent samples were collected for analysis of the challenge chemical(s), TDS, pH, temperature, total chlorine, and turbidity. If a system did not resume operation after sample collection, the additional volume necessary to resume operation was dispensed from each system. 9. The units were then allowed to operate until the shut-off valves activated, and then rest under pressure for at least eight hours. After the rest period, the faucets were fully opened, and the first draw out of each faucet was collected for single challenge chemical and TDS analysis. After collection of the first draw water, the rest of the contents of each storage tank were collected into suitable containers, and three samples were collected from each volume for triplicate challenge chemical analysis. Table 3-2 gives a summary of the sampling plan. Table 3-2. Summary of Sampling Plan for RO Membrane Challenges T ... _ , XT i Effluent Sample Numbers Influent Sample Numbers , . (per system) Water Chemistry Challenge Challenge Sample Point Parameters Chemical TDS Chemical TDS Start Up 1 for each parameter 31 3 1 1st Tank Reject Water Samples Start Up 2 (from one system) Half Tank 2 (from one system) Three-fourths Tank 2 (from one system) 3rd Tank Fill 31 3 1 5th Tank Fill 1 for each parameter 31 3 1 Reject Water - Halfway Through 2 (from one system) 5th Tank Fill Post-Rest - First Draw 1 1 Post-Rest-Rest of Tank 3 15 ------- 3.2.5.3 Post-Membrane Carbon Filter Challenge Testing The post-membrane carbon filter in the Purefecta™ is downstream from the storage tank, so it was tested at the flow rate measured at the faucet outlet during the RO membrane conditioning step. Each challenge was 15 hours. The filters were operated on an "on/off operation cycle where the "on" portion was the time required to empty the storage tank when full, and the "off portion of the cycle was the time required to fill the storage tank at 80 psi inlet pressure, as measured during the RO membrane conditioning period. Figure 3-2 shows a pair of carbon filters being tested. The challenge tests were conducted as follows: 1. The proper "on/off cycle parameters were entered into the test station computer. 2. The initial dynamic inlet water pressure was set at 60 ± 3 psi, and filter operation was started using the water described in section 3.2.3.2 with the proper challenge chemical added. The flow rate was adjusted as necessary using a valve downstream of each filter on the effluent line. 3. Influent and effluent samples were collected for challenge chemical analysis immediately after operation began. All effluent samples were collected during the last half of the "on" portion of the operation cycle, so that the dwell water was flushed out prior to sample collection. All challenge chemical samples were collected and analyzed in triplicate. The sample volumes were those required to obtain the triplicate samples. 4. Single influent samples were also collected for analysis of pH, TDS, temperature, TOC, total chlorine, and turbidity whenever challenge chemical samples were collected. 5. After 7.5 and 15 hours of operation, second and third sets of influent and effluent samples were collected for challenge chemical analysis. The flow of challenge water through the filters was started manually if they were not in the "on" portion of the operation cycle. Table 3-3 gives a summary of the sampling schedule. Figure 3-2. Post-Membrane Carbon Filters Installed at a Test Station. 16 ------- Table 3-3. Summary of Sampling Plan for Post-Membrane Carbon Filter Challenges Influent Water Challenge Chemical Challenge Chemical Chemistry Sample Influent Sample Effluent Sample Sample Point Numbers Numbers Numbers Start Up 7.5 Hours 15 Hours 1 for each parameter 1 for each parameter 1 for each parameter -> 5 -> 5 3 -> 5 -> 5 3 3.3 Analytical Methods 3.3.1 Water Quality Analytical Methods The following are the analytical methods used during verification testing. All analyses followed procedures detailed in NSF's Standard Operating Procedures (SOPs). The reporting limits, and the acceptable precision and accuracy for each parameter are shown in Table 3-4. • pH - All pH measurements were made with an Orion Model SA 720 meter. The meter was operated according to the manufacturer's instructions, which are based on Standard Methods method 4500-H+. • Temperature - Water temperature was measured using an Omega model HH11 digital thermometer. Table 3-4. Parameter pH TDS (conductivity) TDS (gravimetric) TOC Total Chlorine Turbidity Aldicarb Benzene Cadmium Carbofuran Cesium Chloroform Dicrotophos Dichlorvos Fenamiphos Mercury Mevinphos Oxamyl Strontium Strychnine QC Limits and Method Reporting Limit NA 2 mg/L 5 mg/L 0.1 mg/L 0.05 mg/L 0.1 NTU l.Ong/L 0.5 |J,g/L 0.3 |j,g/L lUg/L 1 |J.g/L 0.5 |J,g/L 10 ng/L 0.2 |J.g/L 2|ag/L 0.2 |J,g/L 0.4 |ag/L l.Ojig/L 2 M-g/L 5^g/L Reporting Limits for Acceptable Precision (RPD) 10% 10% 10% 10% 10% 10% 20% 20% 20% 20% 20% 20% RSD < 30% RSD < 30% RSD < 30% 20% RSD < 30% 20% 20% 20% Analyses Acceptable Accuracy (% recovery) 90-110% 80-120% 90-110% 80-120% 90-110% 95-105% 70-130% 80-120% 70-130% 70-130% 70-130% 80-120% 70-130% 70-130% 70-130% 70-130% 70-130% 70-130% 70-130% 80-120% 17 ------- • TDS (by conductivity) - TDS for the TDS reduction system check test was measured through conductivity according to Standard Method 2510 using a Fisher Scientific Traceable™ Conductivity Meter. • TDS (gravimetrically) - The TDS in the carbon filter conditioning and challenge water was measured gravimetrically. The method used was an adaptation of USEPA Methods 160.3 and 160.4. An appropriate amount of sample was placed in a pre-weighed evaporating dish. The sample was evaporated and dried at 103-105 °C to a constant weight. The dish was then weighed again to determine the total solids weight. • Total Chlorine - Total chlorine was measured according to Standard Method 4500-C1 G with a Hach Model DR/2010 spectrophotometer using AccuVac vials. 3.3.2 Challenge Chemical Analytical Methods The following are the analytical methods used during verification testing. All analyses followed procedures detailed in NSF SOPs. The reporting limits, and the acceptable precision and accuracy for each parameter are shown in Table 3-4. • Aldicarb, Carbofuran, and Oxamyl were measured by high pressure liquid chromatography (HPLC) according to USEPA Method 531.1 or 531.2. • Dichlorvos, Dicrotophos, Fenamiphos, and Mevinphos were measured by gas chromatography/mass spectrometry (GC/MS) according to USEPA Method 525.2. • Cadmium, Chromium, Mercury, and Strontium were measured by Inductively Coupled Plasma - Mass Spectrometry (ICP-MS) according to USEPA Method 200.8. • Benzene and Chloroform were measured by purge and trap capillary gas chromatography according to USEPA Method 502.2. • There is no standard analytical method for strychnine. NSF developed a method to measure it using reverse phase HPLC with ultraviolet lamp detection. 18 ------- Chapter 4 Results and Discussion 4.1 RO Membrane Conditioning As discussed in section 3.2.4.1, the RO membrane test units were conditioned for seven days prior to the chemical challenges. The units were conditioned simultaneously at four different test stations, with six units plumbed to each station. All of the water quality parameters in section 3.2.3.1 were maintained within the allowable ranges. The individual data values for these parameters can be found in Table A-l of Appendix A. 4.1.1 RO Membrane Test System Operation Data As described in section 3.2.4.1, the storage tank fill times, first liter dispense times, and times to dispense the entire tanks were measured and recorded for the three systems whose tanks filled the fastest. The first liter flow rates were calculated for each of the three systems from the first liter dispense times. The results are given below in Table 4-1. This data was used to determine the operation parameters for the carbon filter challenges. Table 4-1. RO Membrane Tank Fill Time 1st Liter Time Unit (minutes) (seconds) 16 3 18 69 70 75 20 22 21 System Operation Data 1st Liter Flow Rate (gpm) Tank Dispense Time 0.79 0.72 0.75 4 4 4 min., 44 sec. min., 44 sec. min., 36 sec. 4.2 Post-Membrane Carbon Filter Conditioning As described in section 3.2.4.2, the post-membrane carbon filters were conditioned with water containing 300 ± 90 |ag/L of chloroform until 250 gallons of the conditioning water had passed through them. Eight filters were conditioned first, and then another six were conditioned later. Influent and effluent samples were collected for analysis at start-up, after approximately 125 gallons, and after approximately 250 gallons. All effluent samples were non-detect (< 0.5 ng/L) for chloroform, except one for Unit 11. This unit had an effluent chloroform level of 6.4 ng/L in the 250-gallon sample. The influent water data are given in Table A-2 of Appendix A. 4.3 TDS Reduction System Performance Check After the RO membranes were conditioned, all underwent the TDS reduction test described in section 3.2.5.1. The maximum effluent TDS level measured was 30 mg/L, corresponding to greater than 96% reduction of TDS for all units. The average TDS reduction for the Purefecta™ , as measured during certification testing, was 89.3%, so test units were representative of expected 19 ------- membrane performance. The TDS reduction data for each RO membrane system can be found in Table A-3 of Appendix A. 4.4 RO Membrane Chemical Challenges The RO membrane challenges were conducted according to the procedure in section 3.2.5.2. The tank-fill times were approximately 70 minutes, so the systems were in operation for approximately seven hours per challenge. After each tank-fill period, 2 to 3.5 liters were drawn out of the storage tanks for sample analysis and to disengage the automatic shut-off valve. 4.4.1 Inorganic Chemicals Challenge The inorganic chemicals challenge data are shown in Table 4-2. Each challenge chemical data point is the arithmetic mean of the triplicate sample analyses, except for the post-rest first liter draws, which were only single samples. All individual sample values constituting the triplicate analyses are presented in Table A-4 of Appendix A. As discussed in section 3.1.1, the challenge water also contained TDS at a target concentration of 750 mg/L to serve as an RO membrane integrity check. The TDS reduction data are also presented in Table 4-2. The TDS data points are from single sample analyses. The challenge water chemistry data are presented in Table A-6 of Appendix A. Table 4-2. RO Membrane Inorganic Chemicals Reduction Data Cadmium Cesium Mercury Strontium TDS Sample (ng/L) (ng/L) (ng/L) (ng/L) (mg/L) Start-up Influent Start-up Effluent, Unit 1 Start-up Effluent, Unit 2 3rd Tank Influent 3rd Tank Effluent, Unit 1 3rd Tank Effluent, Unit 2 5th Tank Influent 5th Tank Effluent, Unit 1 5th Tank Effluent, Unit 2 Post-Rest 1st Draw, Unit 1 Post-Rest 1st Draw, Unit 2 Post-Rest 2nd Sample, Unit 1 Post-Rest 2nd Sample, Unit 2 Mean Influent Mean Effluent, Unit 1 Mean Effluent, Unit 2 Percent Reduction, Unit 1 Percent Reduction, Unit 2 Overall Mean Effluent Overall Percent Reduction Units Tested 1000 0.6 1.8 1000 1.8 2.5 1000 1.7 2.3 1.8 2.5 1.8 2.4 1000 1.5 2.3 >99 >99 1.9 >99 1,2 1000 15 31 1000 40 50 1100 37 48 38 52 37 50 1000 33 46 97 95 40 96 1,2 1100 190 460 1100 740 760 1100 810 840 730 760 750 750 1100 640 710 42 35 680 38 1,2 840 2 2 850 2 3 860 2 2 2 3 2 2 850 2 2 >99 >99 2 >99 1,2 740 30 32 760 760(1) 36 750 29 34 34 34 NA NA 750 31 34 96 95 33 96 (1) Sample result not included in mean effluent and percent reduction calculations, see section 5.7.4.3 for further discussion. 20 ------- The RO membrane performed very well against cadmium, cesium, and strontium, removing 96% of the cesium, and more than 99% of the cadmium and strontium. The RO membrane did not perform well against the mercury challenge, but this was expected. There are no POU RO systems certified by NSF for mercury reduction because mercury is not well removed by RO membranes using the test water specified in NSF/ANSI Standard 58. The reject water sample data are given in Table 4-3. The values presented are the arithmetic means of the duplicate sample analyses, except where indicated. The individual sample results are presented in Table A-7 of Appendix A. The chemical levels are somewhat higher than expected, given that the Purefecta™ has an efficiency rating (as defined by NSF/ANSI Standard 58) of 26%. The efficiency is the amount of influent water that is delivered as permeate. This efficiency means the reject water should have approximately 25% more of the challenge chemical than the influent water, assuming 100% rejection by the membrane. The efficiency will be higher at the start of unit operation, since the storage tank is empty, and thus is not supplying any back-pressure. However, by halfway and three-fourths of the way through the tank-fill period, the measured reject water challenge chemical levels are still greater than 25% above the influent levels, indicating that the test units achieved greater than 26% efficiency. Table 4-3. Inorganic Chemicals Challenge Reject Water Data Sample Cadmium (|~ig/L) Cesium (|~ig/L) Mercury (|~ig/L) Strontium (|J.g/L) Start-Up 2100 2000 1200 1/2 Through 1st Tank 1800 1700 1300 3/4 Through 1st Tank 1700(1) 1700(1) 1200 1/2 Through 5th Tank 1500 1500 1200 Unit Sampled 1 1 1 (1) Result is from only one of the duplicate analyses, due to analytical errors. 1700 1400 1400(1) 1200 1 4.4.2 Organic Chemical Challenges The organic chemical challenge data are shown below in Table 4-4. Each data point is the arithmetic mean of the triplicate sample analyses, except for the post-rest first draw samples, which were only single samples. All individual sample values constituting the triplicate analyses are presented in Table A-5 in Appendix A. The water chemistry data for these challenges are presented in Table A-6 in Appendix A. As discussed in section 3.1.1, the challenge water also contained TDS to serve as a membrane integrity check. The TDS reduction data are presented in Table 4-5. The reject water data are shown in Table 4-6. The values presented are the arithmetic means of the duplicate sample analyses, except where indicated. The individual sample results are presented in Table A-8 of Appendix A. The RO membrane removed all chemicals but chloroform by 96% or more. At start-up, the membranes removed greater than 99% of the chloroform, but the effluent levels rose from sample point to sample point after that. The maximum effluent was 310 |~ig/L, from the unit 2 21 ------- Table 4-4. RO Membrane Organic Chemical Challenge Aldicarb Benzene Carbofuran Chloroform Dichlorvos Dicrotophos Fenamiphos Sample (ug/L) (jig/L) (ug/L) (ug/L) (ug/L) (ug/L) (ug/L) Start-Up Influent Start-Up Effluent, Unit 1 Start-Up Effluent, Unit 2 3rd Tank Influent 3rd Tank Effluent, Unit 1 3rd Tank Effluent, Unit 2 5th Tank Influent 5th Tank Effluent, Unit 1 5th Tank Effluent, Unit 2 Post-Rest 1st Draw, Unit 1 Post-Rest 1st Draw, Unit 2 Post-Rest 2nd Sample, Unit 1 Post-Rest 2nd Sample, Unit 2 Mean Influent Mean Effluent, Unit 1 Mean Effluent, Unit 2 Percent Reduction, Unit 1 Percent Reduction, Unit 2 Overall Mean Effluent Overall Percent Reduction Units Tested 960 4 5 920 7 8 980 8 8 8 8 8 8 950 7 7 >99 >99 7 >99 1,2 1100 ND (0.5) 0.7 1200 20 20 1100 67 78 96 85 61 48 1100 49 46 96 96 48 96 3,4 900 ND(1) ND(1) 980 7 5 980 7 6 7 6 7 6 950 6 5 >99 >99 6 >99 5,6 1100 0.7 1.3 1200 71 110 1100 230 320 260 310 150(" 250 1100 140 200 87 82 170 85 7,8 1000 27 33 1100 26 22 1200 18 16 31 19 20 23 1100 23 23 98 98 23 98 11, 12 1000(l) ND(10) ND(10) 740 ND(10) ND(10) 790 ND(10) ND(10) ND(10) ND(10) ND(10) ND(10) 790 ND(10) ND(10) 99 99 ND(10) 99 9, 10 680 ND(2) ND(2) 900 2 3 650 2 2 2 ND(2) 3 3 740 2 2 >99 >99 2 >99 13, 14 Data Mevinphos (ng/L) 1300 15 13 1300 18 21 1500 19 22 12 21 20 23 1400 17 20 99 99 19 99 15, 16 Oxamyl (Hg/L) 970 3 4 980 5 5 1000 5 5 5 4 5 5 980 5 5 >99 >99 5 >99 17, 18 Strychnine (ng/L) 1100 15 ND(5) 1100 29 ND(5) 1100 32 7 34 6 33 6 1100 29 6 97 >99 18 98 19,20 Note: The detection limit values were used for calculating the mean effluents and percent reductions. (1) Number only the average of two of the triplicate analysis numbers, due to analytical errors. post-rest first draw sample. This corresponds to 72% reduction, using the mean influent for the percent reduction calculation. The rising effluent level trend was also evident for benzene. Both of these substances are volatile, so perhaps volatility played a role in their passage through the membrane. They may have adsorbed onto and diffused through the membrane material, or perhaps they began to break down the membrane. The absorption theory is lent some weight by an examination of the reject water data in Table 4- 6. The average concentrations for benzene and chloroform are lower than for the other chemicals. The reject water concentrations for all other chemicals are above the influent challenge levels, indicating that they did not adsorb onto the membrane or surfaces in contact with the water. The organic chemical challenges reject water data also indicates that the test units operated at greater than 25% efficiency, as discussed above in section 4.4.1. If benzene or chloroform did begin to degrade the integrity of the RO membrane, it is not evident through an examination of the TDS reduction data in Table 4-5, nor is it evident that any other chemicals adversely affected membrane performance. More research would be needed to determine why the amounts of benzene and chloroform passing through the membrane increased through the challenge period. 22 ------- Table 4-5. Aldicarb IDS Sample (mg/L) Start-Up Influent Start-Up Effluent, Unit 1 Start-Up Effluent, Unit 2 3rd 'lank Influent 3rd Tank Effluent, Unit 1 3rd Tank Effluent, Unit 2 5th Tank Influent 5th Tank Effluent, Unit 1 5th Tank Effluent, Unit 2 Post-Rest 1st Draw, Unit 1 Post-Rest 1st Draw, Unit 2 Mean Influent Mean Effluent, Unit 1 Mean Effluent, Unit 2 Percent Reduction, Unit 1 Percent Reduction, Unit 2 860 16 99 840 38 21 730 15 22 18 17 810 22 40 97 95 TDS Reduction Data Benzene Carbofuran Chloroform TDS TDS TDS (mg/L) (mg/L) (mg/L) 730 21 15 750 18 12 750 19 13 20 13 740 20 13 97 98 750 130 23 730 17 100 740 18 20 17 13 740 46 39 94 95 740 22 20 740 16 20 750 16 21 16 22 740 18 21 98 97 for Organic Chemical Challenges Dichlorvos Dicrotophos Fenamiphos Mevinphos Oxamyl TDS TDS TDS TDS TDS (mg/L) (mg/L) (mg/L) (mg/L) (mg/L) 760 19 13 750 16 14 740 17 14 17 14 750 17 14 98 98 750 30 27 750 20 23 760 19 22 20 23 750 22 24 97 97 1000 120 1000(l) 830 570 2700 750m 11 15 II 14 750 11 15 99 98 760 16 14 770 17 15 770 16 15 17 15 770 17 15 98 98 760 14 13 750 13 12 750 14 12 14 12 750 14 12 98 98 Strychnine TDS (mg/L) 760 34 17 750 36 15 750 37 15 38 15 750 36 16 95 98 (1) Sample results not included in mean effluent and percent reduction calculations. See section 5.7.4.3 for further discussion. Table 4-6. Organic Chemical Challenge Reject Water Data Sample Start-Up 1/2 through 1st Tank 3/4 through 1 st tank 1/2 through 5th Tank Unit Sampled Aldicarb (Hg/L) 1800 1500 1500 1400 1 Benzene (Hg/L) 970 1100 1000 970 3 Carbofuran (Hg/L) 1300 1600 1500 1400 5 Chloroform (fig/L) 980 990 970 1000 7 Dichlorvos (Hg/L) 2700 1800 1800 1300 9 Dicrotophos (Hg/L) 1400 1200 1700 880 12 Fenamiphos (Hg/L) 1500 1400 1200 1100 13 Mevinphos (Hg/L) 2000 2100 1800 1400 15 Oxamyl (Hg/L) 1700 1500 1500 1300 17 Strychnine (Hg/L) 1900 1800 1700 1500 19 4.5 Post-Membrane Carbon Filter Challenges Based on the RO membrane challenge results, and the criteria discussed in section 3.1.2, the post-membrane carbon filter was challenged with mercury, benzene, chloroform, dichlorvos, mevinphos, and strychnine. The target challenge levels were the maximum effluent levels measured during the RO membrane challenges. Based on the data in Table 4-2, the carbon filters were operated at 0.8 gpm on an operation cycle where the "on" portion was four minutes and thirty seconds, and the "off portion was one hour and ten minutes. The carbon challenge results are shown below in Table 4-7. Each data point is the arithmetic mean of the triplicate sample analyses, except for the footnoted Mevinphos data points. The lab technician did not use the proper preservative for the Mevinphos samples, and the preservative interfered with the internal standard, making it impossible to determine the amount of Mevinphos in some of the samples. See section 5.7.4.3 for further discussion. All individual sample values constituting the triplicate analyses are presented in Table A-9 in Appendix A. The water chemistry data for these challenges can be found in Table A-10 of Appendix A. The carbon filters reduced all substances by 99% or more, except for strychnine. The percent reduction of strychnine was limited by the detection limit for the chemical. 23 ------- Table 4-7. Post-Membrane Carbon Filter Challenge Data Mercury Benzene Chloroform Dichlorvos Mevinphos Sample (ng/L) (ng/L) (ng/L) (ng/L) (ng/L) Target Influent Level Start-Up Influent Start-Up Effluent, Unit 1 Start-Up Effluent, Unit 2 7.5 Hours Influent 7.5 Hours Effluent, Unit 1 7.5 Hours Effluent, Unit 2 15 Hours Influent 1 5 Hours Effluent, Unit 1 15 Hours Effluent, Unit 2 Mean Influent Mean Effluent, Unit 1 Mean Effluent, Unit 2 Percent Reduction, Unit 1 Percent Reduction, Unit 2 Overall Mean Effluent Overall Percent Reduction 840 830 1.4 1.7 IToo 3.4 3.6 970 2.8 3.0 960 2.6 2.8 >99 >99 2.7 >99 96 83 ND (0.5) ND (0.5) 82 ND (0.5) ND (0.5) 85 ND (0.5) ND (0.5) 83 ND (0.5) ND (0.5) >99 >99 ND (0.5) >99 320 320 ND (0.5) ND (0.5) 300 ND (0.5) ND (0.5) 330 ND (0.5) ND (0.5) 320 ND (0.5) ND (0.5) >99 >99 ND (0.5) >99 33 28 ND (0.2) ND (0.2) 28 ND (0.2) ND (0.2) 31 ND (0.2) NDJ0.2) 29 ND (0.2) ND (0.2) >99 >99 ND (0.2) >99 23 21 ND (0.2)(1) ND (0.2) 19 ND (0.2)(1) ND (0.2)(1) 21 ND (0.2)(2) ND (0.2)(2) 20 ND (0.2) ND (0.2) 99 99 ND (0.2) 99 Strychnine (HS/L) 30 37 ND(5) ND(5) 28 ND(5) ND(5) 21 ND(5) ND(5) 31 ND(5) ND(5) 84 84 ND(5) 84 Note: The detection limit values were used for calculating the mean effluents and percent reductions. (1) Mean calculated from only two sample analyses, see section 5.7.4.3 for further discussion. (2) Only one sample analyzed for data point, see section 5.7.4.3 for further discussion. 4.6 Conclusions The RO membrane was not able to remove more than 90% of the chloroform or mercury challenges. However, the membrane and post-membrane carbon filter challenge data combined shows that the two treatment technologies working in concert within the Purefecta™ system removed 99% or more of all challenge chemicals, except for cesium, which was not used to test the carbon filter. 24 ------- Chapter 5 QA/QC 5.1 Introduction An important aspect of verification testing is the QA/QC procedures and requirements. Careful adherence to the procedures ensured that the data presented in this report was of sound quality, defensible, and representative of the equipment performance. The primary areas of evaluation were representativeness, precision, accuracy, and completeness. Because the ETV was conducted at the NSF testing lab, all laboratory activities were conducted in accordance with the provisions of the NSF International Laboratories Quality Assurance Manual. 5.2 Test Procedure QA/QC NSF testing laboratory staff conducted the tests by following an NSF SOP created specifically for the tests. NSF QA Department Staff performed an informal audit during testing to ensure the proper procedures were followed. All water quality measurements were within the specifications in sections 3.2.3.1 and 3.2.3.2, except for TDS in the RO membrane conditioning and challenge water. Two of the influent TDS samples from the Aldicarb challenge, and two of the influent TDS samples from the Fenamiphos challenge were above the allowable upper limit of 825 mg/L (see Table 4-7). However, the high TDS levels were not significant deviations. It is unlikely that these deviations affected the chemical challenge tests in any way. 5.3 Sample Handling All samples analyzed by the NSF Chemistry Laboratory were labeled with unique ID numbers. These ID numbers appear on the NSF laboratory reports for the tests. All samples were analyzed within allowable hold times. 5.4 Analytical Methods QA/QC The calibrations of all analytical instruments, and the analyses of all parameters complied with the QA/QC provisions of the NSF International Laboratories Quality Assurance Manual. The NSF QA/QC requirements are all compliant with those given in the USEPA Method or Standard Method for the parameter. Also, every analytical instrument has an NSF SOP governing its use. 25 ------- 5.5 Documentation All laboratory activities were documented using specially prepared laboratory bench sheets and NSF laboratory reports. Data from the bench sheets and laboratory reports were entered into Microsoft Excel spreadsheets. These spreadsheets were used to calculate average influents and effluents, and percent reductions for each challenge chemical. One hundred percent of the data entered into the spreadsheets was checked by a reviewer to confirm all data and calculations were correct. 5.6 Data Review NSF QA/QC staff reviewed the raw data records for compliance with QA/QC requirements. NSF ETV program staff checked 100% of the data in the NSF laboratory reports against the laboratory bench sheets. 5.7 Data Quality Indicators The quality of data generated for this ETV can be established through four indicators of data quality: representativeness, accuracy, precision, and completeness. 5.7.1 Representativeness Representativeness refers to the degree to which the data accurately and precisely represent the conditions or characteristics of the parameter represented by the data, or the expected performance of the RO system under normal use conditions. Representativeness was ensured by consistent execution of the test protocol for each challenge chemical, including timing of sample collection, sampling procedures, and sample preservation. Representativeness was also ensured by using each analytical method at its optimum capability to provide results that represent the most accurate and precise measurement it is capable of achieving. 5.7.2 Accuracy Accuracy was quantified as the percent recovery of the parameter in a sample of known quantity. Accuracy was measured through use of both matrix spikes of a known quantity, and certified standards during calibration of an instrument. The following equation was used to calculate percent recovery: Percent Recovery = 100 x [(Xknown - XmeaSLlred)/Xknown] where: Xkaown = known concentration of the measured parameter Xmeasured = measured concentration of parameter The accuracy of the benchtop chlorine, pH, TDS, and turbidity meters were checked daily during the calibration procedures using certified check standards. For samples analyzed in batches (gravimetric TDS, TOC, all challenge chemicals), certified QC standards and/or matrix spikes were run with each batch. 26 ------- The percent recoveries of all matrix spikes and standards were within the allowable limits for all analytical methods. 5.7.3 Precision Precision refers to the degree of mutual agreement among individual measurements and provides an estimate of random error. One sample per batch was analyzed in duplicate for the gravimetric TDS, TOC and challenge chemical analyses. Duplicate drinking water samples were analyzed as part of the daily calibration process for the benchtop chlorine, pH, TDS, and turbidity meters. Precision of the duplicate analyses was measured by use of the following equation to calculate relative percent deviation (RPD): RPD = x200 Sl+S2 where: Sl = sample analysis result; and S2 - sample duplicate analysis result. All RPDs were within NSF's established allowable limits for each parameter. 5.7.4 Completeness Completeness is the proportion of valid, acceptable data generated using each method as compared to the requirements of the test/QA plan. The completeness objective for data generated during verification testing is based on the number of samples collected and analyzed for each parameter and/or method. Table 5-1. Completeness Requirements Number of Samples per Parameter and/or Method Percent Completeness (MO 80% 11-50 90% > 50 95% Completeness is defined as follows for all measurements: %C = (V/T)X100 where: %C = percent completeness; V = number of measurements judged valid; and T = total number of measurements. 27 ------- 5.7.4.1 Number of Systems Tested Twenty units were tested, as called for in the test/QA plan, giving a completeness measurement of 100% for this category. 5.7.4.2 Water Chemistry Measurements One hundred percent of the planned samples were collected and reported for every parameter but TOC. During the carbon filter mercury challenge one TOC sample was missed. A total of 18 TOC samples were to be collected during the carbon filter challenges, plus six more during the carbon filter conditioning periods. The missed sample gives a completeness of 96%. 5.7.4.3 Challenge Chemicals Five TDS samples were reported, but not included in the percent reduction calculations in Table 4-3 and Table 4-6 because there was likely a sampling error associated with the sample. The samples were the third tank unit 1 effluent sample from the inorganic chemicals challenge, and the start-up and third tank effluent TDS samples from both test units for the RO membrane fenamiphos challenge. These five TDS analyses were not included in the percent reduction calculations because they were all much higher than expected. The inorganic chemicals challenge third tank unit 1 effluent sample had a reported level of 760 mg/L, the same as the influent. The unit 1 TDS levels from the start-up and fifth tank samples were 30 mg/L and 29 mg/L, respectively, indicating that the unit was functioning properly, and that the third tank sample was an aberration. Of the fenamiphos challenge effluent TDS samples in question, one result was equal to the influent (unit 2 start-up), while the other three were over 100 mg/L. The effluent fenamiphos levels from the same start-up and third tank sample points were at or near the detection limit of 2 Hg/L, which indicates that the test units were functioning properly. Also, the fifth tank and post- rest effluent TDS sample results were all less than 15 mg/L. It is highly unlikely that the test units improved in performance over a matter of hours such that the effluent TDS levels dropped over tenfold. It is likely that there was a sampling error, or the samples were somehow contaminated. Discarding these five samples gives a completeness of 96% for this parameter. One reject water sample from the inorganic chemicals challenge was likely subject to a dilution error, so the results were not used to calculate the mean reject water levels in Table 4-4. The reported cadmium, cesium, and strontium results for the "3/4 through 1st tank" sample are only the second duplicate sample. The first duplicate sample results were approximately twice as high as the second sample. The same sample was analyzed separately for mercury, and the first and second samples both gave results of 1200 |~ig/L. If the amount of the chemicals in the first sample was actually twice that of the second sample, it is likely that it also would have been evident in the mercury analysis. A total of eight reject water samples were collected during the inorganic chemicals challenge. The one sample whose results were discarded gives a completeness of 87.5% for the reject water samples. 28 ------- Two challenge chemical data points in Table 4-5 are the mean of only two of the triplicate sample analyses, due to likely dilution errors during analysis. The unit 1 post-rest second sample from the chloroform reduction challenge is reported as 150 ng/L. This number is the mean of 120 |~ig/L and 180 |~ig/L. The third number triplicate analysis number is more than four times higher than the mean of 150 ng/L, at 690 ng/L. The start-up influent for dichlorvos is reported as 1000 |~ig/L. The two triplicate analyses constituting this mean are both 1000 |~ig/L. The third triplicate number not used is 540 ng/L, which is approximately half of the other two numbers. Both the 690 |ag/L and 540 |ag/L are reported in the appendix in Table A-5. The one sample result discarded for each chemical out of the 35 samples collected (excluding the reject water samples), gives a completeness percentage of 97%. The samples for mevinphos analysis from that chemical's carbon filter challenge were improperly preserved with sodium thiosulfate instead of sodium sulfite. The sodium thiosulfate gave high sulfur levels in the samples, which caused a suppression of the internal standard added during the sample extraction process. This suppression in turn caused an exaggeration of the recovery standard during sample analysis, which made it impossible to accurately determine the Mevinphos levels in some of the effluent samples. Six samples were affected, out of a total of 63 that were collected during both the RO membrane and carbon filter mevinphos challenges. This corresponds to a completeness of 90% for mevinphos. 29 ------- Chapter 6 References APHA, AWWA and WEF (1998). Standard Methods for Examination of Water and Wastewater. 20th ed. Washington, D.C. APHA. NSF International (2005). NSF/ANSI53 - 2005, Drinking water treatment units - health effects. Ann Arbor, MI, NSF International. NSF International (2005). NSF/ANSI 58 - 2005, Reverse osmosis drinking water treatment systems. Ann Arbor, MI, NSF International. Slovak, Robert (2000). A Practical Application Manual for Residential, Point of Use Reverse Osmosis Systems. Lisle, IL, Water Quality Association USEPA (2004). Water Security Research and Technical Support Action Plan. EPA/600/R- 04/063. 30 ------- Appendix A Conditioning and Chemical Challenges Data Tables 31 ------- Table A-l. RO Membrane Conditioning Water Chemistry Data Sample Group A Influent pH Temperature (°C) Total Chlorine (mg/L) TDS (mg/L) Turbidity (NTU) Group B Influent PH Temperature (°C) Total Chlorine (mg/L) TDS (mg/L) Turbidity (NTU) Group C Influent pH Temperature (°C) Total Chlorine (mg/L) TDS (mg/L) Turbidity (NTU) Group D Influent PH Temperature (°C) Total Chlorine (mg/L) TDS (mg/L) Turbidity (NTU) Day 1 7.1 25 ND (0.05) 750 0.1 7.3 25 ND (0.05) 770 ND(O.l) 7.4 24 ND (0.05) 770 ND(O.l) 7.5 25 ND (0.05) 770 ND(O.l) Day 2 7.4 24 ND (0.05) 770 ND(O.l) 7.2 25 ND (0.05) 750 ND(O.l) 7.8 25 ND (0.05) 750 0.1 7.1 24 ND (0.05) 760 0.1 Day 3 7.5 25 ND (0.05) 760 0.1 7.4 25 ND (0.05) 740 0.1 7.5 25 ND (0.05) 740 ND(O.l) 7.2 25 ND (0.05) 730 ND(O.l) Day 4 7.4 25 ND (0.05) 750 0.1 7.4 25 ND (0.05) 740 0.1 7.4 25 ND (0.05) 740 0.1 7.3 25 ND (0.05) 750 0.1 Day 5 7.4 25 ND (0.05) 740 0.1 7.5 25 ND (0.05) 740 ND(O.l) 7.4 24 ND (0.05) 740 ND(O.l) 7.4 24 ND (0.05) 730 ND(O.l) Day 6 7.6 25 ND (0.05) 770 ND(O.l) 7.4 24 ND (0.05) 770 ND(O.l) 7.4 24 ND (0.05) 760 ND(O.l) 7.5 24 ND (0.05) 750 ND(O.l) Day 7 7.2 25 ND (0.05) 760 ND(O.l) 7.1 24 ND (0.05) 750 ND(O.l) 7.1 24 ND (0.05) 740 ND(O.l) 7.2 24 ND (0.05) 740 ND(O.l) Table A-2. Post-Membrane Carbon Filter Conditioning Influent Water Data Total Chloroform Temperature Chlorine Total Organic Sample Point (ng/L) pH (°C) (mg/L) Carbon (mg/L) Turbidity (NTU) Group 1 Start-up 125 gallons 250 gallons Group 2 Start-up 125 gallons 250 gallons 360 310 290 370 350 350 7.5 7.6 7.6 7.4 7.3 7.4 21 21 22 21 21 21 2.5 2.3 2.8 2.8 2.4 2.2 2.5 2.5 2.4 2.0 2.2 2.1 ND(O.l) ND(O.l) 0.2 ND(O.l) ND(O.l) ND(O.l) 32 ------- Table A-3. RO Membrane TDS reduction System Check Data Sample Group 1 Influent Unit 1 Unit 2 Unit 9 Unit 10 Unit 23 Unit 24 Group 2 Influent Unit 3 Unit 4 Units Unite Unit? UnitS Group 3 Influent Unit 1 1 Unit 12 Unit 13 Unit 14 Unit 15 Unit 16 Group 4 Influent Unit 17 Unit 18 Unit 19 Unit 20 Unit 21 Unit 22 TDS Total Influent Effluent Temperature Chlorine Turbidity TDS Sample 1 pH (°C) (mg/L) (NTU) (mg/L) (mg/L) 7.4 25 ND(0.05) ND (0.1) 760 16 16 19 18 14 15 7.4 25 ND(0.05) ND (0.1) 750 16 12 14 11 16 18 7.3 25 ND(0.05) ND(O.l) 750 17 13 11 14 15 14 7.4 25 ND(0.05) ND(O.l) 760 13 12 31 15 14 15 TDS Effluent Sample 2 (mg/L) 16 16 19 18 14 16 16 11 14 11 16 18 17 13 12 15 16 15 13 12 31 15 15 15 TDS Effluent Sample 3 (mg/L) 16 16 19 18 14 16 16 12 14 11 16 18 17 13 11 14 16 15 13 12 31 15 15 15 Percent Reduction 98% 98% 98% 98% 98% 98% 98% 99% 99% 99% 98% 98% 98% 98% 99% 98% 98% 98% 98% 98% 96% 98% 98% 98% 33 ------- Table A-4. RO Membrane Inorganic Chemicals Challenge Data Sample Start-up Influent Triplicate Sample 1 Triplicate Sample 2 Triplicate Sample 3 Mean Start-up Effluent, Unit 1 Triplicate Sample 1 Triplicate Sample 2 Triplicate Sample 3 Mean Start-up Effluent, Unit 2 Triplicate Sample 1 Triplicate Sample 2 Triplicate Sample 3 Mean 3rd Tank Influent Triplicate Sample 1 Triplicate Sample 2 Triplicate Sample 3 Mean 3rd Tank Effluent, Unit 1 Triplicate Sample 1 Triplicate Sample 2 Triplicate Sample 3 Mean 3rd Tank Effluent, Unit 2 Triplicate Sample 1 Triplicate Sample 2 Triplicate Sample 3 Mean 5th Tank Influent Triplicate Sample 1 Triplicate Sample 2 Triplicate Sample 3 Mean 5th Tank Effluent, Unit 1 Triplicate Sample 1 Triplicate Sample 2 Triplicate Sample 3 Mean 5th Tank Effluent, Unit 2 Triplicate Sample 1 Triplicate Sample 2 Triplicate Sample 3 Mean Post-Rest 1st Draw, Unit 1 Post-Rest 1st Draw, Unit 2 Post-Rest 2nd Sample, Unit 1 Triplicate Sample 1 Triplicate Sample 2 Triplicate Sample 3 Mean Post-Rest 2nd Sample, Unit 2 Triplicate Sample 1 Triplicate Sample 2 Triplicate Sample 3 Mean Cadmium (Hg/L) 970 1000 1100 1000 0.6 0.6 0.6 0.6 1.8 1.7 1.9 1.8 1000 1000 1000 1000 1.8 1.7 1.8 1.8 2.6 2.5 2.4 2.5 1000 1000 1000 1000 1.7 1.7 1.6 1.7 2.3 2.3 2.4 2.3 1.8 2.5 1.8 1.8 1.7 1.8 2.4 2.4 2.4 2.4 Cesium (fig/L) 1000 1000 1100 1000 15 15 15 15 32 29 31 31 1100 1000 1000 1000 40 40 40 40 51 49 51 50 1100 1100 1000 1100 37 37 37 37 48 48 49 48 38 52 37 38 36 37 50 49 51 50 Mercury (fig/L) 1100 1100 1100 1100 190 200 190 190 450 490 450 460 1100 1100 1100 1100 710 730 780 740 780 710 780 757 1100 1100 1100 1100 840 780 800 810 840 870 820 840 730 760 750 740 750 750 740 750 760 750 Strontium (fig/L) 800 840 890 840 2 2 2 2 2 2 2 2 840 850 850 850 2 2 2 2 3 3 3 3 860 860 850 860 2 2 2 2 2 2 2 2 2 3 2 2 2 2 2 2 2 2 34 ------- Table A-5. RO Membrane Organic Chemical Challenge Data Aldicarb Benzene Carbofuran Sample (ug/L) (iig/L) (ug/L) Start- up Influent Triplicate Sample 1 Triplicate Sample 2 Triplicate Sample 3 Mean Start-up Effluent, Unit 1 Triplicate Sample 1 Triplicate Sample 2 Triplicate Sample 3 Mean Start-up Effluent, Unit 2 Triplicate Sample 1 Triplicate Sample 2 Triplicate Sample 3 Mean 3rd Tank Influent Triplicate Sample 1 Triplicate Sample 2 Triplicate Sample 3 Mean 3rd Tank Effluent, Unit 1 Triplicate Sample 1 Triplicate Sample 2 Triplicate Sample 3 Mean 3rd Tank Effluent, Unit 2 Triplicate Sample 1 Triplicate Sample 2 Triplicate Sample 3 Mean 5th Tank Influent Triplicate Sample 1 Triplicate Sample 2 Triplicate Sample 3 Mean 5th Tank Effluent, Unit 1 Triplicate Sample 1 Triplicate Sample 2 Triplicate Sample 3 Mean 5th Tank Effluent, Unit 2 Triplicate Sample 1 Triplicate Sample 2 Triplicate Sample 3 Mean Post-Rest 1st Draw, Unit 1 Post-Rest 1st Draw, Unit 2 Post-Rest 2nd Sample, Unit 1 Triplicate Sample 1 Triplicate Sample 2 Triplicate Sample 3 Mean Post-Rest 2nd Sample, Unit 2 Triplicate Sample 1 Triplicate Sample 2 Triplicate Sample 3 Mean 940 960 970 960 4 4 4 4 5 5 5 5 920 980 960 920 7 7 7 7 8 8 8 8 1000 990 960 980 8 8 8 8 8 8 8 8 8 8 8 7 8 8 8 8 8 8 (1) Sample results not included in 1000 1100 1200 1100 ND (0.5) ND (0.5) ND (0.5) ND (0.5) 0.7 0.7 0.7 0.7 1200 1200 1200 1200 17 20 22 20 20 21 20 20 1100 980 1200 1100 68 70 64 67 69 83 82 78 96 85 68 57 58 61 54 49 42 48 880 890 920 900 ND(1) ND(1) ND(1) ND(1) ND(1) ND(1) ND(1) ND(1) 970 970 990 980 7 7 7 7 5 5 5 5 990 980 970 980 7 7 7 7 6 6 6 6 7 6 7 7 7 7 6 6 6 6 Chloroform Dicrotophos Dichlorvos (ug/L) (ug/L) (ug/L) 1000 1100 1100 1100 0.7 0.7 0.8 0.7 1.4 1.3 1.1 1.3 1200 1100 1200 1200 67 80 65 71 100 120 110 110 1100 1200 1100 1100 230 220 230 230 340 310 300 320 260 310 690'" 120 180 150 290 250 220 250 mean calculations. See section 5.7 1000 1000 540(l) 1000 ND(10) ND(10) ND(10) ND (10) ND(10) ND(10) ND(10) ND (10) 780 830 620 740 ND(10) ND(10) ND(10) ND (10) ND (10) ND(10) ND(10) ND (10) 800 800 770 790 ND(10) ND(10) ND(10) ND (10) ND(10) ND(10) ND(10) ND (10) ND(10) ND(10) ND(10) ND(10) ND(10) ND (10) ND(10) ND(10) ND(10) ND (10) .4.3 for discussion. 1200 890 1000 1000 27 28 25 27 33 35 32 33 1100 1100 1100 1100 27 25 25 26 23 22 20 22 1200 1200 1100 1200 17 16 22 18 14 15 18 16 31 19 17 21 23 20 20 23 27 23 Fenamiphos (Ug/L) 600 620 820 680 ND(2) ND(2) ND(2) ND(2) ND(2) ND(2) ND(2) ND(2) 1000 820 880 900 2 2 2 2 3 ND(2) 3 3 640 720 590 650 3 2 ND(2) 2 ND(2) ND(2) 2 2 2 ND(2) 4 3 3 3 3 3 3 3 Mevinphos (Hg/L) 1400 1200 1300 1300 14 16 14 15 9.2 16 13 13 1500 1100 1200 1300 15 16 22 18 20 20 24 21 1200 1300 2000 1500 19 20 17 19 23 18 24 22 12 21 19 21 21 20 24 21 24 23 Oxamyl (Ug/L) 970 960 970 970 3 3 3 3 4 4 4 4 990 980 980 980 5 5 5 5 5 5 5 5 1000 1000 1000 1000 5 5 5 5 5 5 5 5 5 4 5 5 5 5 5 5 5 5 Strychnine (Hg/L) 1100 1100 1100 1100 14 15 15 15 ND(5) ND(5) ND(5) ND(5) 1100 1100 1100 1100 29 29 30 29 ND(5) ND(5) ND(5) ND(5) 1100 1100 1100 1100 32 32 33 32 7 7 6 7 34 6 32 33 33 33 6 6 6 6 35 ------- Table A-6. RO Membrane Challenge Water Chemistry Data Inorganics Aldicarb Benzene Carbofuran Chloroform Dicrotophos Dichlorvos Fenamiphos Mevinphos Oxamyl Strychnine Sample Challenge Challenge Challenge Challenge Challenge Challenge Challenge Challenge Challenge Challenge Challenge Start-up Influent pH Temperature (°C) Total Chlorine (mg/L) Turbidity (NTU) 5th Tank Influent pH Temperature (°C) Total Chlorine (mg/L) Turbidity (NTU) 6.3 25 ND (0.05) ND(O.l) 6.2 25 ND (0.05) ND(O.l) 7.6 25 ND (0.05) ND(O.l) 7.6 25 ND (0.05) ND(O.l) 7.5 25 ND (0.05) ND(O.l) 7.7 25 ND (0.05) ND(O.l) 7.4 25 ND (0.05) 0.1 7.3 25 ND (0.05) 0.1 7.4 25 ND (0.05) 0.1 7.5 25 ND (0.05) ND(O.l) 7.8 25 ND (0.05) ND(O.l) 7.4 25 ND (0.05) ND(O.l) 7.3 25 ND (0.05) ND(O.l) 7.1 25 ND (0.05) ND(O.l) 7.7 25 ND (0.05) ND(O.l) 7.4 25 ND (0.05) ND(O.l) 7.8 25 ND (0.05) ND(O.l) 7.6 25 ND (0.05) ND(O.l) 7.6 25 ND (0.05) ND(O.l) 7.6 25 ND (0.05) ND(O.l) 7.7 25 ND (0.05) ND(O.l) 7.4 25 ND (0.05) ND(O.l) Table A-7. RO Membrane Inorganic Chemicals Challenge Reject Water Data Sample Cadmium (lig/L) Cesium (lig/L) Mercury (lig/L) Strontium (lig/L) Start-up Duplicate Sample 1 2000 2000 1200 1700 Duplicate Sample 2 2100 2000 1200 1700 Mean 2100 2000 1200 1700 1/2 Through First Tank Duplicate Sample 1 1800 1700 1200 1400 Duplicate Sample 2 1700 1700 1300 1400 Mean 1800 1700 1300 1400 3/4 Through First Tank Duplicate Sample 1 3400(1) 36001" 1200 3300(1) Duplicate Sample 2 1700 1700 1200 1400 Mean 1700 1700 1200 1400 1/2 Through 5th Tank Duplicate Sample 1 1400 1400 1200 1100 Duplicate Sample 2 1500 1500 1200 1200 Mean 1500 1500 1200 1200 (1) Sample results not included in mean calculations. See section 5.7.4.3 for discussion. Table A-8. RO Membrane Organic Chemical Challenge Reject Water Data Aldicarb Benzene Carbofuran Chloroform Dicrotophos Dichlorvos Fenamiphos Mevinphos Oxamyl Strychnine Sample (ug/L) (ug/L) (ug/L) (ug/L) (ug/L) (ug/L) (ug/L) (ug/L) (ug/L) (ug/L) Start-up Duplicate Sample 1 1800 1000 1300 990 1900 2600 1300 1900 1700 1900 Duplicate Sample 2 1800 940 1300 970 850 2700 1700 2000 1700 1900 Mean 1800 970 1300 980 1400 2700 1500 2000 1700 1900 1/2 Through First Tank Duplicate Sample 1 1400 1100 1600 1000 840 1700 1300 2300 1500 1800 Duplicate Sample 2 1500 1100 1600 980 1600 1800 1400 1800 1500 1800 Mean 1500 1100 1600 990 1200 1800 1400 2100 1500 1800 3/4 Through First Tank Duplicate Sample 1 1400 890 1500 940 1500 1900 1100 1800 1500 1700 Duplicate Sample 2 1500 1100 1400 1000 1800 1600 1300 1700 1500 1700 Mean 1500 1000 1500 970 1700 1800 1200 1800 1500 1700 1/2 Through 5th Tank Duplicate Sample 1 Duplicate Sample 2 Mean 1400 1400 1400 980 960 970 1400 1300 1400 1100 970 1000 940 810 880 1200 1400 1300 930 1300 1100 1500 1300 1400 1300 1300 1300 1400 1500 1500 36 ------- Table A-9. Post-Membrane Carbon Filter Challenge Data Sample Target Influent Level Start-up Influent Triplicate Sample 1 Triplicate Sample 2 Triplicate Sample 3 Mean Start-up Effluent, Unit 1 Triplicate Sample 1 Triplicate Sample 2 Triplicate Sample 3 Mean Start-up Effluent, Unit 2 Triplicate Sample 1 Triplicate Sample 2 Triplicate Sample 3 Mean 7.5 Hours Influent Triplicate Sample 1 Triplicate Sample 2 Triplicate Sample 3 Mean 7.5 Hours Effluent, Unit 1 Triplicate Sample 1 Triplicate Sample 2 Triplicate Sample 3 Mean 7.5 Hours Effluent, Unit 2 Triplicate Sample 1 Triplicate Sample 2 Triplicate Sample 3 Mean 15 Hours Influent Triplicate Sample 1 Triplicate Sample 2 Triplicate Sample 3 Mean 15 Hours Effluent, Unit 1 Triplicate Sample 1 Triplicate Sample 2 Triplicate Sample 3 Mean 15 Hours Effluent, Unit 2 Triplicate Sample 1 Triplicate Sample 2 Triplicate Sample 3 Mean X - No analysis result due Mercury (fig/L) 840 860 840 800 830 1.5 1.4 1.3 1.4 1.8 1.6 1.6 1.7 1100 1000 1100 1100 3.4 3.4 3.5 3.4 3.6 3.6 3.6 3.6 960 990 960 970 2.7 2.9 2.9 2.8 3.1 3.0 3.0 3.0 to sampling error. Benzene (Hg/L) 96 76 87 87 83 ND (0.5) ND (0.5) ND (0.5) ND (0.5) ND (0.5) ND (0.5) ND (0.5) ND (0.5) 83 79 83 82 ND (0.5) ND (0.5) ND (0.5) ND (0.5) ND (0.5) ND (0.5) ND (0.5) ND (0.5) 78 85 93 85 ND (0.5) ND (0.5) ND (0.5) ND (0.5) ND (0.5) ND (0.5) ND (0.5) ND (0.5) See section 5 Chloroform (Hg/L) 320 330 310 330 320 ND (0.5) ND (0.5) ND (0.5) ND (0.5) ND (0.5) ND (0.5) ND (0.5) ND (0.5) 310 290 300 300 ND (0.5) ND (0.5) ND (0.5) ND (0.5) ND (0.5) ND (0.5) ND (0.5) ND (0.5) 330 340 330 330 ND (0.5) ND (0.5) ND (0.5) ND (0.5) ND (0.5) ND (0.5) ND (0.5) ND (0.5) Dichlorvos (Hg/L) 33 26 29 28 28 ND (0.2) ND (0.2) ND (0.2) ND (0.2) ND (0.2) ND (0.2) ND (0.2) ND (0.2) 28 30 27 28 ND (0.2) ND (0.2) ND (0.2) ND (0.2) ND (0.2) ND (0.2) ND (0.2) ND (0.2) 31 31 30 31 ND (0.2) ND (0.2) ND (0.2) ND (0.2) ND (0.2) ND (0.2) ND (0.2) ND (0.2) Mevinphos (Hg/L) 23 21 21 20 21 ND(0.2) X ND(0.2) ND (0.2) ND(0.2) ND(0.2) ND(0.2) ND (0.2) 19 18 20 19 X ND(0.2) ND(0.2) ND (0.2) ND(0.2) X ND(0.2) ND (0.2) 21 22 21 21 ND(0.2) X X ND (0.2) X ND(0.2) ND(0.2) ND (0.2) Strychnine (Hg/L) 30 37 37 37 37 ND(5) ND(5) ND(5) ND(5) ND(5) ND(5) ND(5) ND(5) 27 28 28 28 ND(5) ND(5) ND(5) ND(5) ND(5) ND(5) ND(5) ND(5) 27 27 27 27 ND(5) ND(5) ND(5) ND(5) ND(5) ND(5) ND(5) ND(5) .7.4.3 for further discussion 37 ------- Table A-10. Post-Membrane Carbon Filter Challenge Water Chemistry Data Sample Mercury Challenge Benzene Challenge Chloroform Challenge Dichlorvos Challenge Mevinphos Challenge Strychnine Challenge Start-up Influent pH Temperature (°C) Total Chlorine (mg/L) TOC (mg/L) TDS (mg/L) Turbidity (NTU) 7.5 20 2.5 2.4 310 0.1 7.3 20 2.6 2.0 290 ND(O.l) 7.4 20 2.3 2.5 320 ND(O.l) 7.3 20 2.6 2.4 290 ND(O.l) 7.4 20 2.7 2.4 290 ND(O.l) 7.2 21 2.3 2.1 320 0.1 7.5 Hour Influent 15 pH Temperature (°C) Total Chlorine (mg/L) TOC (mg/L) TDS (mg/L) Turbidity (NTU) Hour Influent PH Temperature (°C) Total Chlorine (mg/L) TOC (mg/L) TDS (mg/L) Turbidity (NTU) 7.4 20 2.2 # 290 ND(O.l) 7.5 21 2.6 2.2 200 ND(O.l) 7.3 20 2.1 1.9 310 ND(O.l) 7.3 20 2.4 2.0 290 0.1 7.4 20 2.7 2.2 280 ND(O.l) 7.5 21 2.5 2.2 300 ND(O.l) 7.3 20 2.8 2.3 280 0.1 7.3 20 2.7 2.4 290 ND(O.l) 7.5 20 1.8 2.3 300 0.2 7.3 20 2.1 2.6 300 0.3 7.4 20 2.5 2.1 310 0.1 7.3 21 2.4 2.1 300 0.1 # Technician missed sample collection 38 ------- |