RESOLUTION OF SEDIMENT CHEMISTRY QUALITY ASSURANCE CODES If results for these QC samples did not fall within certain pre-established control limits, the analysis of a batch of samples was not considered acceptable. These and other quality control issues are coded in four data qualifier codes (QA_CODE) or "flags" used in the Louisianian Province sediment chemistry data file: CH-A CODE The "CH-A" code indicates that an analyte was not detected. When the "CH-A" code is used, the concentration field is left blank and the method detection limit for the analyte in that particular sample is reported under DETLIMIT. CH-B CODE It is sometimes possible for a laboratory to detect an analyte and report its concentration at a level which is below the calculated method detection limit for the sample. In these situations, the analyst is confident that the analyte was present in the sample, but there is a high degree of uncertainty in the reported concentration. The "CH-B" code is used to flag reported values which are below the calculated method detection limit for the sample. Such values are considered estimates only and should be used with discretion. CH-C The CH-C code indicates that the laboratory experienced minor deficiencies meeting the QC requirements, but the overall data quality is judged to be reliable for EMAP assessments. CH-I CODE Some analytes are difficult to quantify because they co-elute with other closely related analytes. This phenomenon is called "matrix interference". When this occurs, the suspect analyte(s) are given a "CH-I" code and concentration is left blank. CH-M CODE Samples werer lost or destroyed. CH-X CODE In favor of expediency, a laboratory may elect to cease reporting some of the analytes. EMAP protocol only requires that the laboratory analyze a given list of chemicals; when they go beyond this list and report additional chemicals, we include them in our data. The "CH-X" code indicates that an analyte has been excluded from a given set of data. Only "unflagged" or CH-E coded values are considered valid and useful for most assessment purposes. CH-Z CODE Some of the analytes listed represent the sum of concentrations of similar analytes (e.g., PCB_TOT is the sum of the concentrations of all PCB congeners). In the event that the concentrations for all of the individual analytes included in the sum are non-detects (have CH-A code) the sum is missing. This is not technically a non-detect, but a sum of non-detects, hence the CH-Z code.